Device Identification
Alerts and Alarms
Please reference the User Manual for complete information and instructions regarding alerts and alarms.
Alerts
Alerts are indicated on the
Display in flashing yellow.
Description
No Breath
No breath is detected for 45 seconds; the device enters Auto Pulse mode
Low Purity
Oxygen level has dropped below 85%
Low Battery
Remaining battery run time has reached six minutes
Unauthorized Battery
A non-O2 Concepts battery is detected
Low External Power
Voltage to the device is below 18 volts
Alarms
Alarms are indicated on
the Display in flashing red.
Description
No Breath
After 15 minutes in Auto Pulse mode, with no breath detected, device will
shut down
Low Purity
Oxygen level has dropped below 72%
Low Battery
Battery power is depleted, device will shut down
Invalid Flow
Pulse bolus volume is out of valid range
Unauthorized Battery
A non-O2 Concepts battery is detected and no other power source is
available
High External Power
Voltage to the device is above 26 volts
Low External Power
Voltage to the device is below 17 volts
No External Power
Power has been removed from the device
Over Temperature
Pump or enclosure temperature has exceeded limits
Invalid Motor Temperature
Motor temperature has exceeded limits
Invalid Box Temperature
Internal box temperature has exceeded limits