background image

Page 2 of 7 

800 -1063 Rev: A 

Device Settings 

Pulse dose settings 1-5 deliver the following volumes (mL) at 20 breaths per minute. 

Setting 

Setting 

Setting 

Setting 

Setting 

8.0 

16.0 

24.0 

32.0 

40.0 

                                          

 

External Power 

 

 

 

 

   

 

 

                                                                        

 

 

     

 

 

                                  

   

 

 

 

 

 

Installing and Removing Battery 

 

 

 

 

 

  

 
 
 
 

To install the battery, align the battery with the 
back of the bottom panel of the device and slide 
until the battery clicks into place. 

To remove the battery, depress button and slide 
the battery toward you. 

 

To use the AC power supply, connect the 
power cord to the AC power converter 
(brick). Connect the power supply to a wall 
outlet first and then to the power connection 
port of the Freedom. The external power 
source icon will be displayed on the control 
panel.  

The DC power supply powers the Freedom 

from a vehicle’s 12

/15-volt DC outlet in all 

settings. To use the DC power supply, connect 
the power cord to the DC power converter 
(brick). Connect the power supply to a DC 
power outlet first, then to the power 
connection port of the Freedom. The external 
power source icon will be displayed on the 
control panel. 

Both the AC and DC power supplies will 
charge the battery at all settings. 

AC Power Supply 

DC Power Supply 

Power Connection Port 

Summary of Contents for 301-0003

Page 1: ... OxlifeFreedom ProviderOrientation Guide O2 CONCEPTS COM 877 867 4008 800 1063 REV A ...

Page 2: ...re Identification 1 Control Panel 2 LCD Display 3 Oxygen Outlet Cannula Port 4 Exhaust Port 5 Rechargeable Battery 6 External Power Port 7 Air Inlet Ports Air Inlet Filters Understanding the Control Panel 8 Battery Status Battery Charging Indicator 9 Menu Button 10 Power Button 11 Cell Signal Strength 12 Airplane Mode 13 Cell Connected 14 GPS Signal 15 Alert Alarm Indication 16 Breath Detection 17...

Page 3: ...e power cord to the AC power converter brick Connect the power supply to a wall outlet first and then to the power connection port of the Freedom The external power source icon will be displayed on the control panel The DC power supply powers the Freedom from a vehicle s 12 15 volt DC outlet in all settings To use the DC power supply connect the power cord to the DC power converter brick Connect t...

Page 4: ... can be accessed using the device serial number on the O2 Concepts DNA portal at www o2dnaportal com Air Travel The Freedom is an approved portable oxygen concentrator as indicated by the requisite FAA statement on the rear label Please note that many air carriers require prior notification usually at least 48 hours for passengers traveling with portable oxygen concentrators Not all air carriers p...

Page 5: ...g properly check the geographic coordinates near the bottom of the screen indicated by the yellow box in the image below The GPS will display the current latitude and longitude of the unit These coordinates can be verified using a web mapping application such as Google Maps 7 If the GPS does not display the Freedom s correct location confirm the unit has a line of sight to the sky This may require...

Page 6: ...ny maintenance on your Freedom other than routine maintenance described below Any such attempt will void the warranty Air Inlet Filter Cleaning Routine weekly cleaning of the air inlet filter and periodic cleaning of the exterior of the Freedom are the only routine maintenance 1 Remove the Air Inlet Filters 2 Wash the filters by running under warm tap water using a mild detergent 3 Rinse thoroughl...

Page 7: ...ucted above Per manufacturer s instructions disinfect exterior and battery using either Briallianize Plastic Cleaner with SofKloth Oxycide Ecolab or equivalent Cannula Specification The recommended maximum cannula length is seven 7 feet Carry Bag The Freedom carry bag provides protection and allows you to easily take your device with you for your daily activities Always use your Freedom in the sup...

Page 8: ...ay in flashing red Description No Breath After 15 minutes in Auto Pulse mode with no breath detected device will shut down Low Purity Oxygen level has dropped below 72 Low Battery Battery power is depleted device will shut down Invalid Flow Pulse bolus volume is out of valid range Unauthorized Battery A non O2 Concepts battery is detected and no other power source is available High External Power ...

Page 9: ... O2 CONCEPTS COM 877 867 4008 ...

Reviews: