Operating Theory 

The Mineral Mizer utilizes several modern water technologies to cost-effectively purify your tap water supply 
into buffered low-mineral water for dental unit bottles, or mineral-free, autoclave-grade water. 

 

Pre-Filter 

- The yellow dot pre-filter contains a combination sediment & carbon block filter. Its function is 

to remove suspended particles, chlorine and chemicals from the feed water, reducing the incidence of 
reverse osmosis (RO) membrane clogging and chemical degradation. The pre-filter element must be 
replaced every twelve months or earlier depending on the quality of the feed water. 

 

Reverse Osmosis (RO) Module 

- The RO module contains a semi-permeable membrane. Its function is to 

divide the RO feed water into separate permeate (low mineral) and brine (high mineral) streams. This is 
accomplished by applying hydraulic pressure greater than the osmotic pressure in water containing

 

dissolved solids, effectively allowing water molecules to pass through the membrane while rejecting most 
other molecules. The productivity of the RO membrane is measured in 2 ways: 

 

water efficiency – the percentage of RO feed water that is converted to permeate water. 

 

rejection rate – the percentage of minerals rejected from the RO feed water. 

Both of these measures are directly affected by the RO feed water pressure and temperature. The life of the 
membrane is determined by measuring the decline in the membrane’s rejection rate over time. Sans this level 
of measurement, the membrane should be replaced every 24 to 36 months. (This could be earlier depending 
upon quality of feed water). 

DI Filters 

- The red dot DI filters contain two types of beads coated with ion exchange resin. One type absorbs 

molecular calcium and magnesium while the other absorbs chlorides. In exchange, the resins release hydrogen 
and hydroxide ions which combine to create additional water molecules. DI filters have a specific capacity to 
absorb minerals so the volume of DI water produced by a 2 filter set varies based on the mineral content of 
the feed water and the rejection rate of the RO membrane. The DI filters must be replaced when exhausted 
(TDS ppm > 5), and at least once annually. 

Servicing your Mineral Mizer 

Schedule 

This schedule is designed for the average potable water supply and should be followed to ensure the proper 
functioning of your Mineral Mizer. 

PART # 

SERVICE INTERVAL 

PRE-

FILTER 

DI FILTER 

RO MEMBRANE 

POST- 

FILTER 

DWS-RPKSD* 

Annual  

 

DWS-LDPF

 

Only if pre-filter clogs between 

annual service events 

 

 

 

DWS-LRDC2 

Only if TDS rises above 5 ppm 

between annual service events 

 

 

 

DWS-ROM-75HE 

3 years (2 years if TDS IN > 150 

ppm) 

 

 

 

 

 

 

 

 

 

 

These filters may be ordered by contacting the dealer through whom you purchased your Mineral Mizer. Feel 
free to contact O-SO Pure at 800-OSO-PURE (800-676-7873) for assistance in ordering from your dealer or by 
emailing your name, practice phone # and desired part # to 

[email protected]

Summary of Contents for DWS-SADI

Page 1: ...al rev 7 DWS SADI Question or comments Contact O SO Pure at 800 OSO PURE or support osopure com Copyright 2016 O SO Pure All rights reserved Page 1 of 16 MINERAL MIZER DEIONIZED WATER SYSTEM INSTALLATION OPERATION SERVICE MANUAL ...

Page 2: ...e Inspection 5 Pay Attention to All Warnings 5 INSTALLATION GUIDE 6 Locate the Mineral Mizer 6 Connect the Mineral Mizer to plumbing 7 OPERATING THEORY 10 SERVICING YOUR MINERAL MIZER 10 Schedule 10 Filter Stocking Strategies 11 Changing Filters 11 Sanitizing the Mineral Mizer 12 Materials required 12 Other requirements 12 Sanitization Procedure 12 TROUBLE SHOOTING GUIDE 13 WARRANTY 14 THIS PAGE I...

Page 3: ...nded Introduction Congratulations on your purchase of the Mineral Mizer autoclave water system The Mineral Mizer produces deionized DI water from tap water allowing you to exit the boiled bottled distilled water hamster wheel When routinely serviced your Mineral Mizer will economically reliably deliver all of your DI water needs for years to come The DI water produced by the Mineral Mizer normally...

Page 4: ... Take inventory of all items supplied to ensure none are missing The box should contain the following Mineral Mizer Head Filters Spanner Wrench Filter Housings Blue 3 8 Tube Feed Water Tube w Press Reg red drain tubing blsck tank Drain Saddle Angle Stop Valve chrome faucet MODEL DESCRIPTION INSTALL IN TANK S DI WATER DISPENSE RO WATER DISPENSE DWS SADI Mineral Mizer SC or MR steel 6 gallon red fau...

Page 5: ...the Tank is usually placed at the back and the Mineral Mizer is positioned on the right side of the cabinet providing maximum viewing exposure to the right center clear filter sumps If this is a Mineral Mizer II make sure that it is placed within the 15 foot range of the dispensing gun assembly Hand tighten the filter sumps onto the Mineral Mizer head before placing it into the cabinet for spatial...

Page 6: ...Valve on the cold water stub plumbing Reconnect the faucet s cold water tube to the top of the Angle Stop Valve Close the Angle Stop Valve and turn on the cold water supply checking for leaks 6 Install the chrome and red faucets Install the chrome red faucets per its included instructions Location options include the sink pop out or through the countertop overhanging the sink IMPORTANT WARNING PLE...

Page 7: ...rresponding dots on the bracket Hand tighten all the Filter Sumps so they are snug Do not use the provided filter wrench to tighten the filter Sumps unless absolutely necessary Place the Mineral Mizer in its service location 9 Drain Tube black After cutting to the appropriate length connect the Drain Tube between the Mineral Mizer s check valve marked To Drain and the Drain Saddle fitting 10 Feed ...

Page 8: ...oth faucets are all CLOSED This will create enough water pressure in the system to see if there are any possible leaks coming from system Tighten the Filter Sumps with the Sump Wrench if necessary Check the tank for leaks at valve as well Production Startup 15 Open the RO Tank valve 16 Turn ON the water supply by opening the Inline Valve on the Feed Water Tube 17 Open Chrome Faucet 18 Allow the Mi...

Page 9: ...f DI filters 21 The Mineral Mizer is now ready for use To check the mineral content slide the Water Quality Monitor s switch to the FEED position Turn the monitor on by pressing its ON OFF switch The display shows the water s mineral content expressed in ppm TDS This number will normally be in the 000 002 range If the number is higher than 004 allow the tank to fill until the display is consistent...

Page 10: ...es are directly affected by the RO feed water pressure and temperature The life of the membrane is determined by measuring the decline in the membrane s rejection rate over time Sans this level of measurement the membrane should be replaced every 24 to 36 months This could be earlier depending upon quality of feed water DI Filters The red dot DI filters contain two types of beads coated with ion e...

Page 11: ...ilters Some possibilities include consumption by staff taking water for home use or new boiled water applications Changing Filters 1 Turn off the water supply by closing the Inline Valve on the Feed Water Tube 2 If this filter change is the result of a high TDS water reading TDS ppm 5 discard the Tank water by opening the Chrome Faucet over the sink until water flow stops Release the valve 3 Close...

Page 12: ...ed Thoroughly wash hands with soap before beginning installation and or sanitization procedure If your hands come into contact with any unsanitary surface while performing this procedure or the installation hands should be rewashed with soap Any surfaces used to work with to set various components on to dry should be cleaned and disinfected Replacement filters membranes and cleaned components shou...

Page 13: ... NO WATER OR NOT ENOUGH WATER Feed water shut off Turn on feed water Storage tank valve shut off Open tank valve Plugged or crimped lines Remove blockage or crimp in lines Pre filter cartridge clogged Replace pre filter cartridge Low feed water pressure Feed water pressure at membrane must be 40 psi LOW FLOW RATE FROM FAUCET DISPENSER Storage tank valve partially closed Open tank valve completely ...

Page 14: ... 800 OSO PURE to arrange shipment IMPORTANT To ensure proper treatment of source water purchaser must replace all filters and the UV lamp if applicable as directed in this manual Failure to service this product as directed voids this Warranty Filters and UV lamps if applicable must be purchased from O SO PURE for ANY and ALL CLAIMS TO BE BACKED BY O SO PURE DATE OF PURCHASE _______________________...

Page 15: ... Mizer Owner s Manual rev 7 DWS SADI Question or comments Contact O SO Pure at 800 OSO PURE or support osopure com Copyright 2016 O SO Pure All rights reserved Page 15 of 16 this page intentionally left blank ...

Page 16: ...rder Rebate will not be applied retroactively o Only one 1 certificate may be redeemed per eligible O So Pure water system Dealer o Dealer and dealer employees are not eligible for referral fee o When submitting your purchase order to O SO Pure include an image of this certificate clearly showing the purchaser information rebate will be sent by O So Pure directly to the purchaser identified above ...

Reviews: