16) SPECIFICATIONS
Description
Rating
Units
Displayed Travel Accuracy
±10% + ±0.3
mm
Operating Temperature
10 – 40 (50 – 104)
o
C (
o
F)
Operating Relative Humidity
30 - 75
%
Storage Temperature
-10 – 60 (14 – 140)
o
C (
o
F)
Storage Relative Humidity
5 – 95 (Non Condensing)
%
Input Power Voltage (Single Phase)
100 – 240
VAC
Input Power Frequency
50 / 60
Hz
Input Power Current (Maximum)
2.3
A
Weight: ERC 3P
4 (9)
kg (lbs)
ERC 3P with Carrying Case
11 (24)
kg (lbs)
Maximum Continuous On Time
10
Minutes
On Time / Off Time
10 / 90
%
Description
Rating
Classification for Shock
Protection
Class II No reliance on protective earth, double insulated
Applied Part Type Shock
& Leakage Current
Type BF (Basic Floating)
Defibrillation Rating
Not Defibrillator Proof
Anesthetic Use
Do NOT use this equipment in the presence of flammable
anesthetics (Neither AP nor APG)
Ingress Protection
IPX1
Method of Sterilization
Non-Sterile
Pollution Degree
Pollution Degree 2 (Office Environment)
Over Voltage Category
Category II