Operator’s Manual ERC 3P
Model: ERC 3P
LC0216-H-2/2022
Page
11
of
45
For Precice & Precice Unyte Nail
Limb
PRECICE
Model
Nail Diameter
Maximum Distance of
Treated Limb to Surface
of IM Canal
Maximum
Patient Weight
8.5 mm, 9.0 mm, 9.5 mm,
10.0 mm, 10.5 mm
13 mm
57 Kg
C
10.7 mm, 11.5 mm, 12.5 mm
16 mm
114 kg
8.5 mm
13 mm
57 Kg
J
10.7 mm, 12.5 mm
16 mm
114 kg
8.5 mm
13 mm
57 Kg
Tibia
Q
10.7 mm
16 mm
57 Kg
8.5 mm, 9.0 mm, 9.5 mm,
10.0 mm, 10.5 mm
45 mm
57 kg
10.7 mm, 11.5 mm
75 mm
114 kg
A-G (except
C), V, X
12.5 mm
90 mm
114 kg
8.5 mm
45 mm
57 kg
10.7 mm
75 mm
114 kg
H, K, U
12.5 mm
90 mm
114 kg
8.5 mm
45 mm
57 kg
Femur
N, M, P
10.7 mm
75 mm
57 kg
165 - 210 mm
pre-distracted
length
25 mm
Non-weight
bearing
Humerus
L, M
8.5mm
225 - 300 mm
pre-distracted length
45 mm
Non-weight
bearing
For Precice Stryde Nail
Limb
PRECICE
STRYDE
Model(s)
Nail
Diameter
Maximum Distance of
Treated Limb Surface to
IM Canal
Maximum
Patient
Weight
10.0
13mm
150 lbs / 69 kg
11.5
16mm
200 lbs / 91 kg
Tibia
C, SJ
13.0
16mm
250 lbs / 114 kg
10.0
70mm
150 lbs / 69 kg
11.5
85mm
200 lbs / 91 kg
Femur
A, B, C,
E, V, X
13.0
100mm
250 lbs / 114 kg
For Precice Bone Transport Nail
Limb
PRECICE
STRYDE™
Model
Nail Diameter
(mm)
Maximum Distance of
Treated Limb Surface
to IM Canal
Max. Patient Weight
Bearing
Use with partially
threaded screws
Max. Patient Weight
Bearing
Use with fully threaded
screws
10.0
19mm
25lbs/11kg
25lbs/11kg
11.5
19mm
190lbs/86kg
125lbs/57kg
Tibia
C, SJ
13.0
19mm
250lbs/114kg
125lbs/57kg
10.0
64mm
25lbs/11kgs
25lbs/11kg
11.5
69mm
190lbs/86kg
125lbs/57kg
Femur
A, B, BT, D, DT,
E, V, X, SE, SB,
SD, SA
13.0
85mm
250lbs/114kg
125lbs/57kg