Proximity fields
test frequency’s and test levels
Test frequency
(MHz)
Band (MHz)
Modulation
Test level (V/m)
Compliance level
(V/m)
385
380-390
Pulse modulation 18
Hz
27
27
450
430-470
Pulse modulation 18
Hz
28
28
710
745
780
704-787
Pulse modulation 217
Hz
9
9
810
870
930
800-960
Pulse modulation 18
Hz
28
28
1720
1845
1970
1700-1990
Pulse modulation 217
Hz
28
28
2450
2400-2570
Pulse modulation 217
Hz
28
28
5240
5500
5785
5100-5800
Pulse modulation 217
Hz
9
9
Separation distances
Nukute Collare maintains basic safety and essential performance when used in the
electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of Nukute Collare can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and Nukute Collare as recommended below, according to the maximum
output power of the communications equipment.
Rated
maximum
output power
of transmitter
(W)
Separation distance according to frequency of transmitter (m)
From
150 kHz to 80 MHz
d = 1.2
√P
From
80 MHz to 800 MHz
d = 1.2
√P
From
800 MHz to 2.7 GHz
d = 2.3
√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23