The Nukute Collare system maintains basic safety and essential performance when used in
the electromagnetic environment specified below. If the essential performance is lost or
degraded it may cause device unintended operation like rebooting and charging failure.
When malfunction occurs remove device from the disturbance area and restarts the device.
If a malfunction cannot be rectified, please contact to customer service. The customer or the
user of the Collare device should ensure that it is used in such an environment.
Guidance and manufacturer’s declaration –
electromagnetic emissions
The Nukute Collare system is intended for use in the electromagnetic environment specified below. The
customer or the user of the Collare system should assure that it is used in such environment.
RF emissions
CISPR 11
Group 1
Nukute Collare uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment
RF emissions
CISPR 11
Class B
Nukute Collare is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Voltage fluctuations and flickering
IEC 61000-3-3
Complies
Harmonic current emissions
IEC 61000-3-2
Complies