CONFIDENTIAL
VERSION
2.0
33 of 38
Guidance and manufacturer’s declaration –
electromagnetic immunity
The Nukute Collare system is intended for use in the electromagnetic environment specified below. The
customer or the user of the Collare system should assure that it is used in such environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±2 kV, ±4 kV, ±8 kV
Contact discharge
±2 kV, ±4 kV, ±8 kV,
±15 kV
Air discharge
±2 kV, ±4 kV, ±8 kV
Contact discharge
±2 kV, ±4 kV, ±8 kV,
±15 kV
Air discharge
Floors should be wood,
concrete or ceramic tile. If floor
is covered with synthetic
material humidity should be at
least 30%
Rated power frequency
(50/60 Hz) magnetic
fields
IEC 61000-4-8
30A/m, 50/60Hz
30A/m, 50/60Hz
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality be that of
a typical commercial or hospital
environment
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV differential
mode
±2 kV common mode
Mains power quality be that of
a typical commercial or
hospital environment
Voltage dips,
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
0% Ut: 0,5 cycle
At 0°, 45°, 90°, 180°,
225°, 270° and 315°
0% Ut: 1 cycle
And
70% Ut: 25/30 cycles
Single phase: at 0°
0% Ut: 250/300 cycle
0% Ut: 0,5 cycle
At 0°, 45°, 90°, 180°,
225°, 270° and 315°
0% Ut: 1 cycle
And
70% Ut: 25/30 cycles
Single phase: at 0°
0% Ut: 250/300 cycle
Mains power quality should be
that of a typical commercial or
hospital environment.