9.0 Electrical/Environmental Requirements
9.1 Electrical
(Supply Voltage Fluctuations Not to /- 10%)
NU-C200V
115V 60Hz 8.5 Amp 1 Phase UL/UL-C Listed
NU-C200R
115V 60Hz
17 Amp 1 Phase UL/UL-C Listed
NU-C200V-E 230V 50Hz
4 Amp 1 Phase CE Certified
NU-C200R-E 230V 50Hz 8.5 Amp 1 Phase CE Certified
9.2 Operational Performance
(for indoor use only)
Environment Temperature Range: 60⁰F - 85⁰F (15⁰C - 30⁰C)
Environment Humidity:
Maximum relative humidity 80% for temperatures up to 31⁰C decreasing linearly to
50% relative humidity at 40⁰C
Environment Altitude:
6562 Feet (2000 meters) above sea level
9.3 Light Exposure
Standard Fluorescent Lighting @ 150 ft. candles (1614 LUX) maximum intensity.
9.4 Installation Category:
2.0
Installation category (overvoltage category) defines the level of transient overvoltage, which the instrument is designed to
withstand safely. It depends on the nature of the electricity supply and its overvoltage protection means. For example, in
CAT II, which is the category used for instruments in installations supplied from a supply comparable to public mains such as
hospital and research laboratories and most industrial laboratories, the expected transient overvoltage is 2500 V for a 230 V
supply and 1500 V for a 120 V supply.
9.5 Pollution Degree:
2.0
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2
assumes that normally only non-conductive pollution such as dust occurs with the exception of occasional conductivity
caused by condensation.
9.6 Chemical Exposure
Chemical exposure should be limited to antibacterial materials used for cleaning and disinfecting.
CHLORINATED AND HALOGEN MATERIALS ARE NOT RECOMMENDED FOR USE ON STAINLESS STEEL SURFACES.
9.7 EMC Performance (classified for light industrial)
Emissions:
EN61326
Immunity:
EN61326
Class A equipment is intended for use in an industrial environment. In the
documentation for the user, a statement shall be included drawing attention to the fact
that there may be potential difficulties in ensuring electromagnetic compatibility in
other environments, due to conducted as well as radiated disturbances.
WARNING
!