OM0232
|
Rev 7 July/2017
Page 13 of 56
NuAire, Inc.
|
2100 Fernbrook Lane
|
Plymouth, MN 55447
|
U.S.A
|
ph: 763.553.1270
|
fx: 763.553.0459
|
tf: 800.328.3352
|
www.nuaire.com
6.1.1 Rotor assembly
1.
Clean and lubricate the motor shaft by depositing a thin film
of the food grade grease supplied.
2.
Place the rotor above the drive shaft.
3.
Slide the rotor on the drive shaft.
4.
A click indicates that the rotor is locked.
5.
Clean and lubricate the pins by depositing a thin film of the
food grease supplied.
6.1.2 Position the buckets
Swing-out rotors are used with round buckets or micro titration buckets.
1.
Place the buckets in the corresponding numbered locations.
6.1.3 First powering
Activate the switch on the bottom right of the control center.
Lid Open Switch
Lid Open Switch
On/Off Switch position
Under Control Center
On/Off Switch position
Under Control Center