General Options
1.
Door Close Delay Time
This value determines the time, in seconds, to increase the door and parameter heater to duty cycle by 10% after
the inner door is closed. Also inhibits gas and RH control for the same period.
2.
Air Inject Time
This value specifies the time, in seconds, for an injection of air into the chamber
3.
Air Inject Cycle Time
This value specifies the amount of time in minutes between each injection of air into the chamber
4.
Digital Output Interval
This parameter specifies the frequency, in minutes that performance and set point data is output to a device such
as a printer that is only capable of receiving data. If the frequency is specified as zero, no data will be output.
5.
Door Ajar Alarm Delay
Determines how long in minutes the glass door can be open before an alarm is declared.
7.
Settings (text located at the top of screen)
Pressing the next button will go to the second page containing the following:
•
The Filter Maintenance button to access:
Resetting the change filter notice
Setting time to reminder (
24 months/
1/48) 0 disables the feature.
View time remaining until current notice is triggered.
•
The Clear Password button that clears the 4 digit password that was input through the Set Password
screen.
•
Change language Button: Pressing this button brings up a screen with the language choices of English
(default), Spanish, French & German. Pressing the button in that screen of the language desired will
change the user screens to that language.
•
Screen Lock button (default Main Screen Unlocked). Pressing this button will change it to “Main screen
Locked” (Lockout = 1). This puts a “Press to unlock” button on the Nuaire Logo in the Main screen. This
feature stops accidental screen changes by brushing the touch screen. You must press and hold (2
seconds) until the message disappears which unlocks the main screen so that you can navigate into; the
System settings, Alarms, Environmental Set Point and Clock Adjustment screens. Pressing the button
again will change the text to “Main Screen Unlocked (Lockout = 0). This removes the “Press to Unlock”
button from the Nuaire Logo and the Main screen goes back to the default condition.
•
Power Loss Mode cancel button. Cancels Power loss mode. Only use this if it is certain that there is no
condensation in the circulation system. See Section 10.2 for further details about the Power Loss Mode.
Summary of Contents for NU-5710
Page 9: ...OM0249 Page 9 of 64 Rev 4 June 2020 ...
Page 13: ...OM0249 Page 13 of 64 Rev 4 June 2020 ...
Page 24: ...OM0249 Page 24 of 64 Rev 4 June 2020 ...
Page 26: ...OM0249 Page 26 of 64 Rev 4 June 2020 ...
Page 58: ...OM0249 Page 58 of 64 Rev 4 June 2020 ...
Page 59: ...OM0249 Page 59 of 64 Rev 4 June 2020 ...
Page 60: ...OM0249 Page 60 of 64 Rev 4 June 2020 ...
Page 61: ...OM0249 Page 61 of 64 Rev 4 June 2020 ...
Page 62: ...OM0249 Page 62 of 64 Rev 4 June 2020 ...