3.  DPA-2FS Differential Amplifier / Filter Module 

 

3.1. 

DPA-2FS Components 

 
The following items are shipped with the DPA-2FS system: 
 

 

Amplifier / Filter module for the EPMS-07 system 

 

User manual 

 

3.2. 

System Description 

 
The  DPA-2FS  differential  amplifier/filter  is  a  plug-in  unit  for  the  npi  EPMS-07  modular 
system. The DPA-2FS is designed to amplify and filter small bio-electrical signals. The input 
voltage range is ±12 V, the inputs are protected up to 15 V. 2 overload-LEDs indicate if the 
linear region of the amplifier is exceeded. All inputs and the output are BNC connectors. 
 
The  input  impedance  is  10  M

.  Thus,  the  module  can  be  used  for  direct  recording  of 

bioelectric  signals,  e.g.  from  peripheral  nerves.  Using  the  differential  input  configuration 
common  mode  signals  and  noise  are  removed  allowing  recordings  of  signals  with  small 
amplitudes. 
 
DC  offsets  can  be  compensated  either  by  the  using  the  OFFSET  control  or  they  can  be 
eliminated using the AC position of the input switches or by using the HIGHPASS filter. 
 
The GAIN range is x1 to x10000 and can be set with two switches (1-2-5-10-20-50-100 rotary 
switch and 1-10-100 range switch). The output signal is passed through the HIGHPASS and 
LOWPASS filter. 
 
Important:  To  avoid  noise  interference  the  respective  toggle  switch  should  be  in  the  zero 
position if one of the inputs is not used. 
 

3.3. 

Signal Flow Diagram 

 
 
The signal is passed through the DPA-2FS as shown below. 
 
 

 

Summary of Contents for DPA-2FS

Page 1: ...FOR THE DPA 2FS DIFFERENTIAL AMPLIFIER FILTER MODULE FOR EPMS SYSTEMS VERSION 2 0 npi 2015 npi electronic GmbH Bauhofring 16 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 support...

Page 2: ...S E 07 Housing 4 2 5 EPMS 03 5 2 6 PWR 03D 5 2 7 System Grounding 6 EPMS 07 EPMS 03 6 EPMS E 07 6 2 8 Technical Data 6 EPMS 07 EPMS E 07 and EPMS H 07 6 EPMS 07 and EPMS H 07 6 EPMS E 07 6 EPMS 03 6 3...

Page 3: ...efore connecting any system to mains Always use a three wire line cord and a mains power plug with a protection contact connected to ground protective earth Before opening the cabinet unplug the instr...

Page 4: ...ems are shipped with the EPMS 07 housing EPMS 07 cabinet with built in power supply Mains cord Fuse 2 A 1 A slow inserted Front covers Figure 1 Left front view of empty EPMS 07 housing In order to avo...

Page 5: ...wer supply unit the power switch and the fuse are located at the rear of the housing see Figure 2 right 2 6 PWR 03D The external power supply PWR 03D is capable of driving up to 3 EPMS E housings Each...

Page 6: ...to the SYSTEM GROUND SIGNAL GROUND on the rear panel of the instrument see Figure 4 The chassis can be linked to PROTECTIVE EARTH by connecting it to the PWR 03D with the supplied 6 pole cable and by...

Page 7: ...nce is 10 M Thus the module can be used for direct recording of bioelectric signals e g from peripheral nerves Using the differential input configuration common mode signals and noise are removed allo...

Page 8: ...DPA 2FS User Manual version 2 0 page 8 3 4 Description of the Front Panel and Operation Figure 5 DPA 2FS front panel view...

Page 9: ...OFFSET potentiometer 10 turn potentiometer for compensating for a DC OFFSET Note The overall OFFSET compensation is the product of the reading of both controls Important Position 5 of the OFFSET poten...

Page 10: ...LEDs indicating that the DPA 2FS exceeds its linear range 10 V 7 OUT connector BNC connector providing the conditioned signal 8 9 IN IN connector BNC connectors for connecting the input signal The in...

Page 11: ...i 15 1641 1645 Huemmeke M Eysel U T Mittmann T 2004 Lesion induced enhancement of LTP in rat visual cortex is mediated by NMDA receptors containing the NR2B subunit J Physiol 559 875 882 Schulz D Hust...

Page 12: ...1 1 V or 10 V set by toggle switch HIGHPASS Filter single pole attenuation 6 dB octave corner frequencies Hz 0 3 5 100 300 400 500 600 800 1k 3k 10k LOWPASS Filter single pole attenuation 6 dB octave...