Set- Up

 

 

Before you use the Bedfont NOxBOX

i

 intelligent Nitric Oxide delivery and monitoring system it is important to follow these steps:

Ensure the device does not require maintenance (monthly) and that the maintenance is not due during the expected duration of the therapy such as:

•  NO sensor high calibration

•  NO

sensor high calibration

•  Pump calibration

•  Vent calibration

Checklist

•  Ensure the device is clean and free from damage

•  Ensure the water trap is empty and has been cleaned from previous use

•  Ensure all single use items from previous patient have been discarded

Equipment  you will need for set-up:

•  A NOxBOxi system test kit (NOXBOX-I TEST)

•  A NOXKIT (inspiratory limb size dependant)

•  Two cylinders of delivery gas

•  A 22mm one-way valve for HFO ventilator circuits (FXS555)

Additional items

•  NOxAIR Environmental monitors

•  Manual bagging kit

NOTE: If you do not have the parts listed above or the device has not been maintained the NOxBOX

i

 may not be suitable for use and you may be unable to 

set-up the device. Failure to follow these procedures may result in harm to to the operator or the patient.

Summary of Contents for NOxBOXi

Page 1: ...tenance is not due during the expected duration of the therapy such as NO sensor high calibration NO2 sensor high calibration Pump calibration Vent calibration Checklist Ensure the device is clean and...

Page 2: ...st is complete follow instructions on screen to set up system Experienced users may choose to skip directly to system tests or skip to dose System test circuit set up NOxFlowTM arrow points in directi...

Page 3: ...regulator surfaces and O rings are damage free Do not use damaged equipment Attach regulator to cylinder Screw firmly by hand Attach supply line to regulator Apply supply line to port on rear of monit...

Page 4: ...run test System test takes approx 5 minutes to run This purges NO2 and ensures correct NOxBOXi operation System test passed Stop O2 flow and disconnect from NOxflow Ready to connect to patient System...

Page 5: ...or with luer port from the NOXKITTM Connect the male luer of the sample line 1 2 3 4 NOxFlowTM is situated in the inspiratory limb before the humidifier The sample line is up to 30cm from patient to t...

Page 6: ...change dose press edit dose The NOxBOXi will then guide you through setting the dose and high low alarms Once confirmed the delivery and monitoring will adjust accordingly Maximum dose 80ppm Minimum...

Page 7: ...er Check the concentration matches the value displayed on screen Check O rings and attach regulator to new cylinder 1 2 5 Perform a leak test Slowly open the cylinder valve then close Observe gauge fo...

Page 8: ...stow on system trolley Remove all single patient use devices from monitor and dispose of according to local regulations Drain water trap and dispose of drainage syringe On completion of last instructi...

Page 9: ...any issue with the NOxBOXi system Alarm Priorities The system alarms are colour coded to help identify priority of detected issue Additionally the two alarm priorities each have an audible warning to...

Page 10: ...the battery may completely discharge and the sensors will lose their calibration bias In the event of power loss connect the unit to mains power and allow 6 hours for the unit to charge before calibr...

Page 11: ...ical guide for flow specifications you may need to increase the ventilator bias flow The NO low alarm may be inappropriately set by user Check NO low alarm value and reduce value if ventilator setting...

Page 12: ...Stagnant gas in manual bag circuit causing NO2 Purge manual bag circuit before connecting to patient see manual bagging procedure The NO2 sensor may require replacing If another system is not readily...

Page 13: ...blocked pinched or occluded Sample monitoring is affected which may compromise deliveryaccuracy and patient safety Check sample line for any pinch crush points from external bodies or blockages that m...

Page 14: ...printed on the NOxFLOW should be pointing towards the patient in the direction of the ventilator flow Check NOxFLOW flow detection lines and connection including O rings are connected and there are n...

Page 15: ...tion Fail Medium One or all of the sensors have failed the zero low calibration Ambient conditions may be affecting the zero sample Check ambient NOxAIR monitor for high levels of NO If high levels ar...

Page 16: ...librations are not performed during therapy to minimise risk to the patient To resolve the system sensors must be calibrated by the appointed Systems Engineer using calibration gases Manual Override M...

Page 17: ...nother system is not readily available and patient requires therapy engage the manual override mode replace the system as soon as practically possible and alert the Service Engineer System Test Fail T...

Page 18: ...30cm between the patient Y Piece 8 and the sample line 7 1 Ventilator 2 Ventilator Inspiratory Port 3 NOxFLOWTM use 22F or 15M to vent tube adaptors 4 Humidifier 5 0 7m 1 3m Corrugated Tubing 15mm or...

Page 19: ...mode selection valve to the position seen on right hand side image 2 A A B B Using on screen calculator access the manual override calculator from Settings Adjust vent flow to best match patient sett...

Page 20: ...8 38 32 28 24 22 20 18 16 15 14 13 12 12 11 10 9 9 9 4 9 8 6 8 3 7 9 7 6 7 4 7 1 6 8 6 6 6 4 6 2 26 5 8 5 7 5 5 5 4 5 2 5 1 5 200 286 167 118 91 74 63 54 48 43 38 35 32 30 28 26 24 23 22 21 20 19 18 1...

Page 21: ...3 13 12 12 12 11 11 10 10 9 8 9 5 9 2 8 9 8 7 8 5 8 2 8 7 8 500 200 133 100 80 67 57 50 44 40 36 33 31 29 27 25 24 22 21 20 19 18 17 17 16 15 15 14 14 13 13 13 12 12 11 11 11 11 10 10 9 8 600 218 150...

Page 22: ...0 5 0 5 0 5 0 4 0 4 0 4 0 4 0 4 0 3 0 3 0 3 0 3 0 3 0 3 0 3 0 3 0 3 0 3 0 3 0 3 0 2 100 17 9 1 6 3 4 8 3 8 3 2 2 8 2 4 2 2 2 1 8 1 6 1 5 1 4 1 3 1 2 1 2 1 1 1 1 0 9 0 9 0 9 0 8 0 8 0 8 0 7 0 7 0 7 0 7...