2.

 

SAFETY INFORMATION 

To avoid possible electric shock or personal injury: 

 

Use the Calibrator only as specified in this manual, or the protection provided 

by the Calibrator might be impaired. 

 

Do not apply more than the rated voltage between the terminals, as marked on 

the Calibrator, or between any terminal to earth.  

 

Make  sure the battery door is closed and latched before operating the 

calibrator. 

 

Remove test leads from the calibrator before opening the battery door. 

 

Inspect the Calibrator before using it. Do not use the Calibrator if it appears 

damaged. 

 

Do not operate the calibrator around explosive gas, vapor, or dust. 

To avoid possible damage the calibrator or system, make sure to use the right 

connector and select the appropriate range before using the instrument for 

measurement or simulation.

 

 

 

 

 

3.

 

MODEL DC80T 

The MODEL DC80T is capable of measuring and generating mV and 

thermocouple signals. The specifications shown on the tables below are valid for 

the period of one year, at ambient temperatures between 18~28 °C. 

INPUT/OUTPUT RANGE

 

RESOLUTION

 

ACCURACY

 

-10 mV to 75 mV 

0.01 mV 

± (0.025 % + 2 counts) 

Measurement and Simulation specifications for Thermocouples: 

Type 

Range 

Resolution 

Accuracy 

Tamb Error (*) 

J  -200 to 1200 °C / -328 to 2192 °F  0.1 °C / ºF 

±

 (0.15 % Span) 

±

 0.5 ºC 

K  -200 to 1370 °C / -328 to 2498 °F  0.1 °C / ºF 

±

 (0.15 % Span) 

±

 0.5 ºC 

-200 to 400 °C / -328 to 752 °F  0.1 °C / ºF 

±

 (0.15 % Span) 

±

 0.5 ºC 

E  -200 to 950 °C / -328 to 1742 °F  0.1 °C / ºF 

±

 (0.15 % Span) 

±

 0.5 ºC 

-20 to 1750 °C / -4 to 3182 °F 

1 °C / ºF 

±

 (1 ºC + 10 µV) 

±

 0.5 ºC 

-20 to 1750 °C / -4 to 3182 °F 

1 °C / ºF 

±

 (1 ºC + 10 µV) 

±

 0.5 ºC 

B  600 to 1800 °C / 1112 to 3272 °F  1 °C / ºF 

±

 (1 ºC + 10 µV) 

±

 0.5 ºC 

N  -250 to 1300 °C / -418 to 2372 °F  0.1 °C / ºF 

±

 (0.3 ºC + 10 µV) 

±

 0.5 ºC 

- (*) Tamb Error is the accuracy of internal ambient temperature sensor 
- Maximal input voltage: 30 Vpp. 

 
 

Summary of Contents for DC80L

Page 1: ...L SPECIFICATIONS 9 5 2 MODEL DC80L FRONT PANEL 9 5 3 PROTECTION 10 5 4 MODEL DC80L OPERATING INSTRUCTIONS 10 6 MAINTENANCE 13 1 INTRODUCTION The Novus Calibrators series DC80 are portable instruments...

Page 2: ...e instrument for measurement or simulation 3 MODEL DC80T The MODEL DC80T is capable of measuring and generating mV and thermocouple signals The specifications shown on the tables below are valid for t...

Page 3: ...to 50 C Shock Random 2 g 5 Hz to 500 Hz Safety 1 meter drop test Power requirements 6 x AAA 1 5 V Battery Size 205 mm x 98 mm x 46 mm Weight 472 g with batteries 3 2 MODEL DC80T FRONT PANEL The keypad...

Page 4: ...large calibration errors Thermocouple or mV measurement 1 Press the power switch to turn on the calibrator 2 Press the INPUT OUTPUT key to select INPUT mode 3 Press the TC TYPE selection key to config...

Page 5: ...target instrument OHM meter of RTD meter connected to the calibrator RTD measurement and simulation specifications Mode Range Accuracy C Excitation current mA Input 4 W Input 2 W 3 W Output Pt10 385...

Page 6: ...mm x 98 mm x 46 mm Weight 472 g including batteries 4 2 MODEL DC80R FRONT PANEL The keypad and display functions are shown below 4 wire input terminal not used for output 2 wire input output terminal...

Page 7: ...ding value RTD Simulation 1 Press the power switch to turn ON the calibrator 2 Press the INPUT OUTPUT key to select OUTPUT mode 3 Press RTD mode key to select the desired RTD type 4 Press the Incremen...

Page 8: ...2 wire transmitter The following tables show the DC80L specifications functions and ranges based on a one year calibration period at ambient temperatures ranging from 18 C to 28 C Counts means units o...

Page 9: ...2 g 5 Hz to 500 Hz Safety 1 meter drop test Power requirement 6 x AAA Size 204 mm L x 99 mm W x 46 mm H Weight 460 g including batteries External power supply 9 Vdc 500 mA 5 2 MODEL DC80L FRONT PANEL...

Page 10: ...calibrator Fuse 2 is probably blown out if the calibrator reads 0 000 in mA input mode even under the presence of a current input signal 5 4 MODEL DC80L OPERATING INSTRUCTIONS DC mV and V measurements...

Page 11: ...egative test lead black to the target equipment 7 For changing the output value use the increment decrement keys and the V mV mode key until the desired value is displayed DC mA measurement a Using ex...

Page 12: ...splay 4 Plug the red test lead into the LOOP jack and the black one to the mA jack 5 Connect the test leads such that the current enters the red lead and returns to the power supply through the black...

Page 13: ...ough the black lead 7 To change the output value press the increment decrement keys and or the mA mA mode key Accessories a pair of Test Leads 6 MAINTENANCE Cleaning Periodically wipe the case with a...

Reviews: