10 

5.3.

 

PROTECTION 

The MODEL DC80L calibrator uses two protection fuses, rated 125 mA / 250 V, 

internally connected in series with the current terminals (2) and (4). 
To avoid personal injury or damage to the calibrator, use only 125 mA / 250 V fast 

fuses. 

Notes on fuses: 

-

 

Fuse 1 is probably blown out when the display flashes the message OL while 

in Voltage output mode with the leads removed from the calibrator. 

-

 

Fuse 2 is probably blown out if the calibrator reads 0.000 in mA input mode, 

even under the presence of a current input signal 

5.4.

 

MODEL DC80L OPERATING INSTRUCTIONS 

DC mV and V measurements 

1.

 

Press the power switch    to 

turn on the Calibrator. 

2.

 

Press the INPUT  /  OUTPUT 

mode key    to select INPUT 

mode. 

3.

 

 Press the V/mV mode key   

to select mV input  . 

4.

 

Plug the red test lead to “V” jack 

  and the black one to the 

COM” jack  . 

5.

 

Connect the test leads 

according to the signal polarity: 

red to the positive polarity, black 

to the negative polarity (GND). 

6.

 

Measurement value is shown on 

the display  . 

 

Summary of Contents for DC80L

Page 1: ...L SPECIFICATIONS 9 5 2 MODEL DC80L FRONT PANEL 9 5 3 PROTECTION 10 5 4 MODEL DC80L OPERATING INSTRUCTIONS 10 6 MAINTENANCE 13 1 INTRODUCTION The Novus Calibrators series DC80 are portable instruments...

Page 2: ...e instrument for measurement or simulation 3 MODEL DC80T The MODEL DC80T is capable of measuring and generating mV and thermocouple signals The specifications shown on the tables below are valid for t...

Page 3: ...to 50 C Shock Random 2 g 5 Hz to 500 Hz Safety 1 meter drop test Power requirements 6 x AAA 1 5 V Battery Size 205 mm x 98 mm x 46 mm Weight 472 g with batteries 3 2 MODEL DC80T FRONT PANEL The keypad...

Page 4: ...large calibration errors Thermocouple or mV measurement 1 Press the power switch to turn on the calibrator 2 Press the INPUT OUTPUT key to select INPUT mode 3 Press the TC TYPE selection key to config...

Page 5: ...target instrument OHM meter of RTD meter connected to the calibrator RTD measurement and simulation specifications Mode Range Accuracy C Excitation current mA Input 4 W Input 2 W 3 W Output Pt10 385...

Page 6: ...mm x 98 mm x 46 mm Weight 472 g including batteries 4 2 MODEL DC80R FRONT PANEL The keypad and display functions are shown below 4 wire input terminal not used for output 2 wire input output terminal...

Page 7: ...ding value RTD Simulation 1 Press the power switch to turn ON the calibrator 2 Press the INPUT OUTPUT key to select OUTPUT mode 3 Press RTD mode key to select the desired RTD type 4 Press the Incremen...

Page 8: ...2 wire transmitter The following tables show the DC80L specifications functions and ranges based on a one year calibration period at ambient temperatures ranging from 18 C to 28 C Counts means units o...

Page 9: ...2 g 5 Hz to 500 Hz Safety 1 meter drop test Power requirement 6 x AAA Size 204 mm L x 99 mm W x 46 mm H Weight 460 g including batteries External power supply 9 Vdc 500 mA 5 2 MODEL DC80L FRONT PANEL...

Page 10: ...calibrator Fuse 2 is probably blown out if the calibrator reads 0 000 in mA input mode even under the presence of a current input signal 5 4 MODEL DC80L OPERATING INSTRUCTIONS DC mV and V measurements...

Page 11: ...egative test lead black to the target equipment 7 For changing the output value use the increment decrement keys and the V mV mode key until the desired value is displayed DC mA measurement a Using ex...

Page 12: ...splay 4 Plug the red test lead into the LOOP jack and the black one to the mA jack 5 Connect the test leads such that the current enters the red lead and returns to the power supply through the black...

Page 13: ...ough the black lead 7 To change the output value press the increment decrement keys and or the mA mA mode key Accessories a pair of Test Leads 6 MAINTENANCE Cleaning Periodically wipe the case with a...

Reviews: