Novis APMBL-L01 User Manual Download Page 16

16

User Guide 

BetterLiving

Troubleshooting

SYMPTOM

SOLUTION

Control unit does not 
operate; no display lights

Check control unit is connected to the mains power supply.

Check for loose power cord connection and ensure main power is 
switched on.

Check the fuse in rear panel of control unit. Replace if necessary.

Check condition of power cord and plug. Check if mains socket is 
faulty.

Pressure Loss indicator 
constantly illuminated; 
mattress is not inflating  
with control unit  
connected and  
switched on

Ensure the main power is turned on and power cord is connected 
to mains and control unit.

Check control unit/mattress air connections are fitted securely, 
and reconnect umbilical cord if loose.

Ensure control unit is turned on.

Check air intake from filter is not blocked by linen/dust. Replace 
with new filter if needed.

Ensure the air cells are free of damage or leaks.

For faster mattress reinflation once the air leak has been closed, press 

MAX INFLATE

 and wait until  

the Pressure Loss indicator extinguishes. Press the MODE button to resume Alternation mode.

Control unit is making 
unusual noise

Ensure control unit is resting against a solid surface

Power Failure

Check control unit is connected to the mains power supply.

Check for loose connection on plug and main power is switched 
on.

Check the fuses in control unit. Replace if necessary.

Check condition of power cord and plug. Check if mains socket is 
faulty.

Patient is sinking or 
‘bottoming out’ whilst lying 
flat on the mattress

The pressure maybe set too low for the patient’s weight – increase 
the pressure setting by cycling to a higher pressure setting on the 
Pressure Setting button.

Check for air leaks in the mattress and air hoses.

   If the problem persists, move patient to an alternate product and contact Novis Healthcare.

Summary of Contents for APMBL-L01

Page 1: ...MATTRESS OVERLAY AlternatingMattressReplacement Overlay U S E R G U I D E MATTRESS REPLACEMENT...

Page 2: ...sional Ifyouhaveanyquestions pleasecontactNovis Healthcareon1300738885 NovisHealthcarehasapolicyofcontinuous productimprovementandreservesthe righttoamendspecificationspresentedin thisguide Informatio...

Page 3: ...osesandCPRrelease Controlunit Powercord Userguide Carrybag Itisrecommendedthatallpacking materialsandmanualsbekeptinthecarry bagprovided foreaseofstorageand or transport Intended Use Thepreventionandt...

Page 4: ...ble pressure injury prevention strategy avoid wearing clothing that may cause areas of localised damage due to creases seams objects in pockets etc Never use sharp objects or electrically heated blank...

Page 5: ...oxygenornitrousoxide Donotuseinthepresenceofsmoking materialsoropenflame airflowingthroughthemattresswillsupport combustion Donotopenthecontrolunit riskof electricalshock Referservicingtoqualified ser...

Page 6: ...a which may risk damage to the mattress replacement BedBase UmbilicalCord FootSymbol Attachtothebedbysecuringtheadjustable straps locatedontheundersideofthemattress baseundereachbedend Onaprofilingbed...

Page 7: ...ddingfromthemattressand placethemattressoverlayontheexisting mattress orsuitablefoamunderlay Printedtopcoverfacingupwardsand umbilicalhosetowardsthebaseofthebed CheckCPRsealingvalveisclosed the turnin...

Page 8: ...clickasconfirmation theconnectorislockedinplace Straighten any twists in the umbilical air hoses Ensure the umbilical air hose is not trapped between the mattress and bed Feedpowercordthroughthecordre...

Page 9: ...lie a person or any weight on the mattress during initial inflation Wheninitialinflationiscomplete the PressureSettingandMaxInflateindicators willextinguish toindicatethesystemis readyforuse Thesyste...

Page 10: ...nderneaththepatientandfeelforadeflatedcellunderthepatient slower spine Theinnerstaticcellwillremaininflated howeveryourhandshouldeasilyslidebetween patientandbase Ifyourhandcanpassunderthepatient then...

Page 11: ...dicatethemodecurrentlyactive SeeModeonpage12formoredetails D Power Pressandholdforaminimumofthreeseconds toturnsystempoweronandoff Green light poweron E Alarm Mute Turnsaudiblealarmofftemporarily Pres...

Page 12: ...to ensureallaircellsareinflated Beforemovingthemattress ensurethecontrol unitisswitchedoffanddisconnectedfrom mainspower Disconnecttheumbilicalairhose fromthecontrolunitbyunpluggingthequick releasecon...

Page 13: ...flateallcells Pressthereleasebuttonsontheumbilicalcord connectorstoreleasetheairhoses Onceairhasbeenreleasedfromthesystem detachthemattressfromthebed Rollthe mattressfromheadendtofootendtoremove addit...

Page 14: ...rol It is recommended that the system is cleaned every two weeks if in constant use Base Cleaning Swabthemattressbaseandcellswithasolutionofsodiumhypochloriteorsimilar upto10 000 ppmavailablechlorine...

Page 15: ...moisture Asoftbristlednylonbrushcanbeusedtogentlycleancrevices Ensure the control unit is disconnected from mains power before cleaning Donotspraydisinfectantdirectlyontothecontrolunit orimmersetheun...

Page 16: ...nen dust Replace withnewfilterifneeded Ensuretheaircellsarefreeofdamageorleaks Forfastermattressreinflationoncetheairleakhasbeenclosed pressMAXINFLATEandwaituntil thePressureLossindicatorextinguishes...

Page 17: ...propriateaccordingthelegislation Please beenvironmentallyresponsibleandcontact yourlocalauthorityonavailableoptionsto recycle thisproductatitsendoflife Service Life Theexpectedservicelifeofacontroluni...

Page 18: ...m canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter wherepisthemaximumoutputpower ratingofthetransmitterinwatts W accordingtothetransmittermanufacturer NOTE1 At 80 MHz and 800 MH...

Page 19: ...0 15MHz 80MHz PortableandmobileRFcommunications equipment shouldbeusednoclosertoanypartofthecontrolunit includingcables thantherecommendedseparation distancecalculatedfromtheequationapplicabletothe fr...

Page 20: ...B 217Hz 0 2 0 3 9 9 745 780 810 800 960 GSM800 900 TETRA800 iDEN 820 CDMA850 LTE Band5 Pulse modulationB 18Hz 2 0 3 28 28 870 930 1 720 1700 1990 GSM1800 CDMA 1900 GSM1900 DECT LTEBand1 3 4 25 UMTS Pu...

Page 21: ...z MAXIMUM CURRENT 0 2 to 0 1A FUSE RATING T2AL250V MIN MAX PRESSURE 20 60mmHg 6mmHg PROTECTION TYPE ClassIITypeBF INGRESS PROTECTION RATING IP21 HEIGHT x WIDTH x DEPTH 195 x 265 x 120mm WEIGHT 2kg MAT...

Page 22: ...ed toitspremisesoroneofitsAuthorisedService Agentsinpersonorfreightprepaidbyyou Noviswillundertakeatitsoption torepairor replace freeofcharge eachproductorpart thereofontheconditionthat Theproductfoun...

Page 23: ...ndthefailure doesnotamounttoamajorfailure Toobtain compensation youwillneedtoprovided documentaryevidenceofthelossordamage sufferedanddocumentaryevidencethatsuch lossordamagewasareasonableforeseeable...

Reviews: