7.7 Customer Interface
Contact Factory for guidance
7.8 RS-422 Compatibility
Optira Series Encoders are RS-
422 compatible. Encoder signals are “sending end terminated”
and customer receiving terminations are not required.
For more details, see the following Tech Note:
Proper Termination of Digital Incremental
Optional RS-422 compliant circuitry for long cable runs in harsh electronic environments is
illustrated below:
7.9 Troubleshooting
Problem
Solution
The LED on the sensor does not light.
•
Confirm power connection is correct.
•
Confirm sensor configuration.
The LED on the sensor displays
steady red.
•
Confirm sensor alignment.
•
Confirm scale cleanliness.
Calibration procedure won’t initiate.
•
Check sensor configuration for LSB.
•
Check connection between Connector and Calibration
board and the sensor.
Calibration procedure won’t complete.
•
Check alignment for Index detection.