Section 7
Electronic Tests
Procedure
32
Model 2001 Service Manual
Rev. 01
RED ADC
5B32-88CB hex
IR ADC
3324-4CB7 hex
14.
Press Quit followed by Yes. Verify a “Monitor performing self test” message is displayed then
is immediately followed by the main monitoring screen.
15.
Press and hold the Contrast key. Verify the display contrast changes from light to dark. Set the
contrast to a normal level.
16.
Press the Event key. Verify an “Event Marked” message is momentarily displayed then a
“Waveform Frozen” message is displayed. Press the Run softkey.
17.
Set the averaging time to 2 seconds. Press the Menu softkey followed by the Avg
and
2s
softkey.
18.
Set the pulse beep to 5. Press the Audio softkey followed by the Pulse softkey. Use the Up /
Down arrow softkeys to set the pulse beep to 5. Press the Run softkey.
19.
Set the TB500B to the settings in the chart. Verify the saturation values are within the specified
tolerance and an audible pulse beep is heard:
Note: Verify a visual (alert bar flashing) and audible alert condition after the saturation is set to
82. Press and hold the audio alert key. Verify the 2 minute silence LED momentarily turns on
then the audio off LED turns on (Blinking) and the audible alert is silenced.
20.
Change the TB500B power switch to OFF. Verify an “Check Sensor Site” error message is
displayed.
21.
Change the TB500B power switch to ON. Verify the error message clears and a saturation and
pulse value are displayed.
22.
Set the TB500B Signal Attenuation to “1”. Verify a “Sensor Off Patient” message is displayed.
23.
Set the TB500B Signal Attenuation to “3”. Verify the error message clears and a saturation and
pulse value are displayed.
24.
Set the TB500B Saturation to “0”. Verify a “Low Signal Strength” message is displayed.
25.
Set the TB500B Saturation to “100”. Verify the error message clears and a saturation and pulse
value are displayed.
26.
Press the TB500B Red Open button. Verify a “Faulty Sensor” message is displayed.
27.
Release the Red Open button. Verify the error message clears and a saturation and pulse value
are displayed.
28.
Press the TB500B Infrared Open button. Verify a “Faulty Sensor” message is displayed.
TB500B Settings
Board Parameters
Saturation
Attenuation
Saturation
Pulse Rate
100
3
98 - 100
60 ± 1
82
3
80 - 84
60 ± 1
62
7
58 - 66
60 ± 1
72
7
68 - 76
60 ± 1
92
7
88 - 96
60 ± 1
Summary of Contents for MARS PO2 TECH 2001
Page 2: ......
Page 66: ...Section 12 Parts Lists 60 Model 2001 Service Manual Rev 01...
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......