8. „STANDBY“ signals operational readiness. Activate the control
unit by pushing „I“. The LED-display „OPERATE“ comes on.
Return to „STANDBY“ by re-pushing of „I“.
9. Now push the mode button to select alternating (LED
– display „ALTER“ lights up) or static mode (LED –
display „STATIC“ lights up) depending on the desired
operating mode,
10. Duration of inflation is 2-5 min.
11. The „Cycle“- button sets the alternating intervals. This
function if only active in the “Alter” mode. Setting is achieved
in 5-minute increments (0-95 Min) by pushing the button.
Push the “Alter” Button to return to „5“.
12. Set patient weight by pushing the -/+ key.
Use setting chart for support.
13. The patient can now be placed onto the
mattress.
14. Now verify or correct the settings by conducting the manual check.
15. Please note that the system pressure might require adjustment when raising
the head rest section.
8. Lock – Function :
In order to prevent unauthorized changes in the default setting, the LAS 500
control unit is equipped with an automatic key lock function. It is automatically
activated after 3 minutes.
Release the key lock function by pushing the button with the key symbol. Keep it
pushed for 5 seconds, until the „Lock“ LED Display goes out.