8.   „STANDBY“ signals operational readiness. Activate the control 

unit  by  pushing  „I“.  The    LED-display  „OPERATE“  comes  on. 
Return to „STANDBY“ by re-pushing of  „I“.  

 
 
 

9.  Now push the mode button  to select alternating (LED 

–  display  „ALTER“  lights  up)  or  static  mode  (LED  – 
display  „STATIC“  lights  up) depending on the desired 
operating mode,  

 

10. Duration of inflation is  2-5 min. 
 
 
11. The  „Cycle“-  button  sets  the  alternating  intervals.  This 

function if only active in the “Alter” mode. Setting is achieved 
in  5-minute  increments  (0-95  Min)  by  pushing  the  button. 
Push the “Alter” Button to return to „5“. 

 
 
 
 
12. Set  patient  weight  by  pushing  the  -/+  key. 

Use setting chart for support.  

 
 

13. The  patient  can  now  be  placed  onto  the 

mattress. 

14. Now verify or correct the settings by conducting  the manual check. 

15. Please  note  that  the  system  pressure  might  require  adjustment  when  raising 

the head rest section.  

 
 

8.  Lock – Function 

In  order  to  prevent  unauthorized  changes  in  the  default  setting,  the  LAS  500 
control  unit  is    equipped  with  an  automatic  key  lock  function.  It  is  automatically 
activated after 3 minutes. 
Release  the  key  lock  function  by  pushing  the  button  with  the  key  symbol.  Keep  it 
pushed for 5 seconds, until the „Lock“ LED Display goes out.

 

 

 

 

 

 

Summary of Contents for LAS 500

Page 1: ...Operating Instructions LAS 500 Low air Loss System...

Page 2: ...tion Page 5 6 8 Lock Function Page 6 9 Backup Function Page 7 10 Static mode Page 7 11 Bariatric Heavy weight function Page 7 12 Care Function Page 7 13 Alarm function Page 7 14 Practical Test Page 8...

Page 3: ...placement system for clinical stationary and ambulatory care taking The system consists of a control unit with digital operating unit and a mattress system with 18 arranged in a transverse manner 2 Se...

Page 4: ...eight observe chart 2 Setting of alternating pressure interval Cycle 3 Alternating mode 4 Static mode 5 Care Mode 6 Start button Operate or Standby 7 display alternating pressure interval Cycle 5 95 i...

Page 5: ...the bed using the fixation hooks The automatically retractable hooks adjust to the foot end of the bed and provide a secure hold Upon removal of the control unit the hooks automatically retract to the...

Page 6: ...ments 0 95 Min by pushing the button Push the Alter Button to return to 5 12 Set patient weight by pushing the key Use setting chart for support 13 The patient can now be placed onto the mattress 14 N...

Page 7: ...tric mattress optional in combination with this function The mode is activated by pushing the BARIATRIC Button The Bariatric LED lights up 12 Use of CARE Function The LAS 500 control unit features a C...

Page 8: ...d once again without difficulty The extracted cell can in re inserted in the opposite sequence 16 CPR Valve The novacare LAS 500 Low Air Loss Therapy System is equipped with a CPR emergency valve It p...

Page 9: ...with RL 93 42 EWG Use and applications are subject to these legal foundations Systems are suitable for reuse However hygiene directives must be observed and flawless operation must be ensured Mainten...

Page 10: ...cribed by the respective manufacturer Guarantee claims become null and void if inappropriate and unlisted preparations are used or are used inappropriately We recommend our cleaning service Please ref...

Page 11: ...al PU Max Flow 1200 ltr min Cycle Adjustable 5 95 Minutes in 5 Minute intervals Surrounding conditions Temperature Application operation 5 50 Transport storage 10 60 Rel humidity Application operation...

Page 12: ...medical care products They are sold by novacare gmbh primarily to medical experts care taking personnel specialized medical dealers users medical facilities etc and put at the disposal of professional...

Page 13: ...intervals prescribed by the manufacturer are adhered to As far as warranty claims are asserted and if during the subsequent inspection it is found that the damages are of wear and tear or are not subj...

Page 14: ...LAS_500_eng page 14 31 Notes...