61
Appendix
7. If in Data Review mode when the test strip is inserted,
the meter immediately switches to test mode. Upon exit-
ing test mode the meter screen goes blank and does not
return to Data Review mode.
8. Battery low icon is displayed in every mode except setup.
9. Once battery level goes below the threshold that trig-
gers the “low battery” warning, it continues to give the
warning until the meter becomes unusable due to low
battery.
10. A failed control solution does NOT lock the user out of
the meter.
11. Beeps that are not disabled when ‘BEEP’ is set to OFF:
The ‘quick triple beep’ warning for HI or LO errors and
Error codes E-0 to E-4.
12. “LO” and “HI” results are omitted from any averages, but
are uploaded.