Interfering Substances
Seventeen (17) interference compounds were studied to
determine if their presence affected the reporting of lactate
results in whole blood. None of the compounds interfered
with the reporting of the Lactate Results at the tested con-
centration range.
Compound
Concentration Range
Tested
Acetaminophen
0-10 mg/dL
Ascorbic Acid
0-10 mg/dL
Bilirubin
0-15 mg/dL
Cholesterol
0-500 mg/dL
Creatinine
0-6 mg/dL
Dopamine
0-10 mg/dL