58
Appendix
Precision
Precision of Nova Lactate Plus Meter was measured with
both venous blood samples and control solution in labora-
tory. The results for the blood samples are shown below:
Lactate Level
1
2
3
4
5
Lactate Mean
(mmol/L)
20 measurements
1.2
3.4
6.8
11.7
16.5
CV%
9.4
4.6
3.3
4.8
3.2
S.D. (mmol/L)
0.11
0.16
0.22
0.56
0.52
Lact Plus Meter PN 62659A84.indd 58
Lact Plus Meter PN 62659A84.indd 58
11/18/19 5:11 PM
11/18/19 5:11 PM