Vacuson 18
Operating Instructions, Vacuson 18, Art. No. 31790, V06/14
15
EN
9.7
Seals and tubes
To insure proper function of the Suction Pump, all tubes and seals must be periodically checked, and
replaced after at most 250 sterilization cycles or five years of operation. Only when in perfect order can
sufficient vacuum be built up.
Defective seals and tubes must be replaced.
1.
Suction tube, PVC, single use, sterile, length 5 x 2 x 4000 mm (optional) -------------------- --------------- Art. No. 6026
2.
Suction tube, silicone, sterilizable, 6 x 2 x 1700 mm (PATIENT) ---------------------------------- 1 unit ------- Art. No. 26014
3.
Connecting tube, silicone, sterilizable, 8 x 3 x 400 mm (VACUUM) ---------------------------- 1 unit -------- Art. No. 4075
4.
Float rubber seal for overflow protection system --------------------------------------------------- 1 unit ------Art. No. XXXXX
5.
Ring seal for secretion jar lid------------------------------------------------------------------------------ 1 unit ------Art. No. XXXXX
6.
Seal (O-Ring) at connection nozzle of control unit for bacteria filter mount ----------------- 2 units ------ Art. No. 4063
7.
Bacteria filter, hydrophobe, PTFE, single use, Ø 64 mm ------------------------------------------- 1 unit -------- Art. No. 4246
2
1
3
4
5
6
7