MD20
NOUVAG AG
2
No. 31842 14/06
1min on /
3min off
1 Product description
1.1 Use and operation
The microprocessor-controlled MD20 is used in implantology and microsurgery. The device is designed
for bone drilling, screwing and sawing. The MD20 includes a hose pump for the cooling of rotating
instruments for the prevention of tissue damage.
1.2 Technical data
Voltage: ............................................................................ switchable: 100 V
∼
/ 115 V
∼
/ 230 V
∼
, 50-60 Hz
Power supply fuse: ...................................................................................................................... 2 x T2A
Power: ............................................................................................................................................160 VA
Application component: ............................................................................................................... Type BF
Class of protection: ........................................................................................................................Class II
Motor rotational speed: ..................................................................................................500 - 40´000 rpm
Max. motor torque: ....................................................................................................................... 10 Ncm
Mass (HxWxD): ........................................................................................................135 x 285 x 260 mm
Net weight: ..........................................................................................................................................6 kg
Printer: ......................................................................................................................................DPT100-S
1.3.
Ambient conditions for operation
Rel. air humidity: ................................. max. 80%
Temperature: ..................................... 10 to 40°C
Air pressure: .............................. 800 to 1060hPa
1.4.
Ambient conditions for transport
and storage
Rel. air humidity:................................. max. 90%
Temperature:........................................0 to 60°C
Air pressure: ............................. 700 to 1060 hPa
1.5.
Device symbols
: Caution: refer to enclosed
documentation
: Application component of type BF
1275 : EU conformity symbol
: Class II equipment
: Certified by Canadian Standards
Association (CSA) for Canada
and USA
: Functional earth terminal
: Autoclavable up to max. 134°C
IP68
: Watertight footpedal
: continuous operation with
intermittent loading
1min on, 3min off
: Old electrical and electronic
equipment must be disposed
separately and may not be
included in regular domestic
waste.
Summary of Contents for md 20
Page 2: ......
Page 4: ......
Page 85: ...NOUVAG AG Anhang Appendix Appendice Appendice Apéndice ...
Page 86: ......