User Manual for Patient Monitor
31
during normal operation. Factory default is
“OFF”.
Notch:
ECG waveform filter, “50 Hz” To do 50hz filtering said ECG waveform, “60 Hz” To do 60hz filtering
said ECG waveform,”OFF”
Do not do filtering,
Factory default i
s “50Hz”.
Grid:
The grids will be displayed on the background if this setting is “ON”. The factory default is ON.
Move the cursor to the upper or lower limits of the alarm settings, and press the Alarm key to turn ON
and OFF the alarm for the setting. Yellow color shows ON status, and gray color show the OFF
status.
4.8.4 NIBP Parameter Settings
SYS Hi/Lo:
Systolic pressure upper and lower alarm limits.
DIA Hi/Lo:
Diastolic pressure upper and lower alarm limits.
MAP Hi/Lo:
Mean arterial pressure upper and lower alarm limits.
PR Hi/Lo:
Pulse rate upper and lower alarm limits.
Unit:
the pressure unit. mmHg or kPa. The factory default is mmHg.
Mode:
The cuff inflation mode, manual or automatic. The factory default is manual. The operator
needs to press the NIBP button to perform NIBP measurement.
>>
:
Next page.
Nibp Cail.:
This option is
for the provision of
factory
calibration
of
the
machine
,
"OFF
",
"
Mode
1"
and
"
Mode
2
".
The
factory
default value: "
Off".
initP.:
initial pressure of nibp measurement.
Figure 4.33 NIBP Settings