User Manual for Patient Monitor
28
4.8 The Menu Setup Screen of the System
4.8.1 Screen Description
During the Mode Selection screen, move the cursor
to the “SETUP” and press the knob. The screen shows
the system setting screen as shown in Figure 4.26.
Figure 4.26 System Settings
To set up the system parameter, rotate the knob and move the cursor to the corresponding button, press
the knob to perform the corresponding system settings.
Press the DEF button will make settings returned to the factory default and will not change the patient
document and the recalled data. Our instructions will cover the functions of each button.
Figure 4.27 System Parameter Settings
Type
: The subject being monitored, this can be chosen among Adult and Pediatric patient.
The factory default is “Adult”
When changing the patient type, the system will perform initiation of the alarm settings, NIBP
settings. Please pay special attention to the patient type before starting the monitoring. It is
strictly forbidden to use Adult type on the pediatric patient. Doing so can cause serious injury.
Mode
: Monitoring mode selection. The “Near real” shows the near real-time waveform, i.e. normal
monitoring state. The DEMO shows the demo waveforms. During the demo state, all the signals
and data are generated from the patient monitor for demo and testing purpose. The factory
default is “Near Real-time”.
Lang
: Language is the setting for the display language. The factory default is English.
Fill
: When the fill setting is ON, the display fills the volume for the SpO2. When it is OFF, the system
displays the line graph. The default is OFF.
Frze
: Freeze controls the waveforms being frozen when the Freeze button is pressed. The options are
“All” and “ECG”. When ECG is selected, the system only freezes the ECG waveform. When “All”
is selected, the system freezes all the waveforms including ECG, SpO2. The factory default is
“ECG”
AlmVol
: The alarm sound volume. The max volume is 7 and min is 0, i.e. no sound. The factory
default is 5.
Key
: keystroke sound. The factory default is ON.
Bee
p: The Pulse rate sound volume. The max volume is 7 and min is 0, i.e. no sound. The factory