Connect Power
NOTES:
• Use only the supplied power cord.
• Ensure that the power source is located near the equipment and is easily accessible.
• Use only on a circuit with a fuse or circuit breaker that is 20 A or less.
1. Connect the power cord to the back of the controller and to your local power source.
2. Connect a 16 AWG (1.3 mm) wire to the chassis grounding screw on the rear of the chassis using a toothed
grounding lug. The wire must have green insulation with a yellow stripe or must be noninsulated (bare).
3. Attach the opposite end of the wire to a permanent earth ground using toothed washers or a toothed lug.
16 AWG
(1.3 mm)
Installation (continued)
ValveMate 9000 Controller
20