Nonin Onyx II 9560 Instructions For Use Manual Download Page 9

9

 

NOTES:

• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 

structures, objects, and people.

The following table details the recommended separation distances between portable and mobile RF communications equipment and this 
device.

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

in meters (m) can be 

estimated using the equation applicable to the frequency of the transmitter, where 

is the maximum output power rating of the transmitter 

in watts (W) according to the transmitter manufacturer. 

NOTES:

• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 

structures, objects, and people.

Table 3: Guidance and Manufacturer’s Declaration—Electromagnetic Immunity

Immunity Test

IEC 60601 Test 

Level

Compliance 

Level

Electromagnetic Environment—Guidance

This device is intended for use in the electromagnetic environment specified below. The customer and/or user of this device should ensure that it is 

used in such an environment.

Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended 
separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended Separation Distance

Conducted RF
IEC 61000-4-6

3 Vrms
150 kHz to 80 MHz

N/A

Radiated RF
IEC 61000-4-3

3 V/m
80 MHz to 2.5 GHz

[3] V/m

80 MHz to 800 MHz 

800 MHz to 2.5 GHz 

Radiated RF per ISO 
9919 clause 36 and
ISO 80601-2-61 clause 
202.6.2.3

20 V/m
80 MHz to 2.5 GHz

[20] V/m

where 

P

 is the maximum output power rating of the transmitter in watts (W) 

according to the transmitter manufacturer and 

is the recommended separation 

distance in meters (m). 

Field strengths from fixed RF transmitters, as determined by an electromagnetic 
site survey

a

, should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following 
symbol: 

a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, 

AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment 
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the 
device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as reorienting or relocating the device.

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [3] V/m.

Table 4: Recommended Separation Distances

This device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Users 

of this device can help prevent electromagnetic interference by maintaining a minimum distance between portable and 

mobile RF communication equipment (transmitters) and the device as recommended below, according to maximum output 

power of the communications equipment.

Separation Distance According to Frequency of Transmitter

Rated Maximum Output 

Power of Transmitter

W

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz

0.01

0.12

0.12

0.23

0.1

0.37

0.37

0.74

1

1.2

1.2

2.3

10

3.7

3.7

7.4

100

12

12

23

d

1.17 P

=

d

1.17 P

=

d

2.33 P

=

d

1.17 P

=

d

1.17 P

=

d

2.33 P

=

Summary of Contents for Onyx II 9560

Page 1: ...tional hemoglobin Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following The device may not work when circulation is reduced Warm or rub th...

Page 2: ...r assistance RF Exposure For body worn operation to maintain compliance with FCC RF exposure guidelines use only accessories that contain no metallic components and provide a separation distance of 15...

Page 3: ...ch new Model 9560 When batteries are low the numeric displays flash once per second Remove batteries if the device will be stored for more than 30 days Replace low batteries as soon as possible using...

Page 4: ...the startup sequence the device begins sensing the pulse indicated by the blinking pulse quality indicator Allow the device to stabilize and observe about 4 seconds of continuous green colored pulse...

Page 5: ...as permanent damage could result 2 Dry with a soft cloth or allow to air dry Ensure that all surfaces are completely dry Testing Summary SpO2 accuracy and low perfusion testing was conducted by Nonin...

Page 6: ...imately 600 spot checks Storage 12 months minimum Classifications per IEC 60601 1 CAN CSA C22 2 No 601 1 UL 60601 1 Degree of Protection Type BF Applied Part Enclosure Degree of Ingress Protection IP3...

Page 7: ...l 9560 found to be within specifications Model 9560 is a precision electronic instrument and must be repaired by trained Nonin personnel only Any sign or evidence of opening the Model 9560 field servi...

Page 8: ...netic environment specified below The customer and or user of this device should ensure that it is used in such an environment Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV cont...

Page 9: ...20 V m 80 MHz to 2 5 GHz 20 V m where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in mete...

Reviews: