Warranty

NONIN MEDICAL, INCORPORATED, (Nonin) warrants to the purchaser, for a period of 4 years from the date of purchase, each Model 9560 
exclusive of the batteries, spring, carrying case, lanyard, and lanyard lock. 

Nonin shall repair or replace any Model 9560 found to be defective in accordance with this warranty, free of charge, for which Nonin has 
been notified by the purchaser by serial number that there is a defect, provided notification occurs within the applicable warranty period. 
This warranty shall be the sole and exclusive remedy by the purchaser hereunder for any Model 9560 delivered to the purchaser which is 
found to be defective in any manner whether such remedies be in contract, tort or by law. 

This warranty excludes cost of delivery to and from Nonin. All repaired units shall be received by the purchaser at Nonin’s place of business. 
Nonin reserves the right to charge a fee for a warranty repair request on any Model 9560 found to be within specifications. 

Model 9560 is a precision electronic instrument and must be repaired by trained Nonin personnel only. Any sign or evidence of opening the 
Model 9560, field service by non-Nonin personnel, tampering, or any kind of misuse of the Model 9560, shall void the warranty. All non-
warranty work shall be done at Nonin’s standard rates and charges in effect at the time of delivery to Nonin.

Nonin Medical, Inc.

13700 1st Avenue North

Plymouth, Minnesota 55441-5443 USA

+1 (763) 553-9968

(800) 356-8874 (USA and Canada) 

Fax: +1 (763) 553-7807

E-mail: [email protected]

Nonin Medical AB

 Fibervägen 2 

 82450 Hudiksvall, Sweden

+46 650 401500 (Europe)

Fax: +46 650 401514

E-mail: [email protected]

www.nonin.com

Antenna:

Type:

L-shaped PWB whip-type antenna

Transmitter:

Bluetooth Compliance:

Version 2.0

Operating Frequency:

2.4 to 2.4835 GHz

Output Power:

<20 dBm

Operating Range:

328 feet/100-meter radius indoors (line of sight when connected to a 
class I device)

Network Topology:

Star

Operation:

Bluetooth Slave

Antenna Type:

Internal

Modulation Type:

Frequency Shift Keying

Frequency Hopping Spread Spectrum:

Band Width:

1 MHz

Summary of Contents for Onyx II 9560

Page 1: ...tional hemoglobin Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following The device may not work when circulation is reduced Warm or rub th...

Page 2: ...r assistance RF Exposure For body worn operation to maintain compliance with FCC RF exposure guidelines use only accessories that contain no metallic components and provide a separation distance of 15...

Page 3: ...ch new Model 9560 When batteries are low the numeric displays flash once per second Remove batteries if the device will be stored for more than 30 days Replace low batteries as soon as possible using...

Page 4: ...the startup sequence the device begins sensing the pulse indicated by the blinking pulse quality indicator Allow the device to stabilize and observe about 4 seconds of continuous green colored pulse...

Page 5: ...as permanent damage could result 2 Dry with a soft cloth or allow to air dry Ensure that all surfaces are completely dry Testing Summary SpO2 accuracy and low perfusion testing was conducted by Nonin...

Page 6: ...imately 600 spot checks Storage 12 months minimum Classifications per IEC 60601 1 CAN CSA C22 2 No 601 1 UL 60601 1 Degree of Protection Type BF Applied Part Enclosure Degree of Ingress Protection IP3...

Page 7: ...l 9560 found to be within specifications Model 9560 is a precision electronic instrument and must be repaired by trained Nonin personnel only Any sign or evidence of opening the Model 9560 field servi...

Page 8: ...netic environment specified below The customer and or user of this device should ensure that it is used in such an environment Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV cont...

Page 9: ...20 V m 80 MHz to 2 5 GHz 20 V m where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in mete...

Reviews: