Specifications

Oxygen Saturation Display Range:

 

0% to 100% SpO

2

Pulse Rate Display Range:

 

18 to 321 beats per minute (BPM)

Declared Accuracy:

The table below shows A

rms

 values measured using the Onyx II 9560 in a clinical study in non-motion conditions.

Pulse Rate Declared Accuracy Range (A

rms

*):

20 to 250 BPM ±3 digits

Low Perfusion Pulse Rate Declared Accuracy Range 
(A

rms

*): 

40 to 240 BPM ±3 digits

Measurement Wavelengths and Output Power**: 

Red:
Infrared:

660 nanometers @ 0.8 mW maximum average
910 nanometers @ 1.2 mW maximum average

Temperature: 

Operating:

23 °F to 104 °F / -5 °C to 40 °C

Storage/Transportation: 

-40 °F to 158 °F / -40 °C to 70 °C***

Humidity: 

Operating:

10% to 95% non-condensing 

Storage/Transportation: 

10% to 95% non-condensing

Altitude: 

 

Operating:

Up to 40,000 feet / 12,192 meters

Hyperbaric Pressure: 

Up to 4 atmospheres

Battery Life:

Operating:

Two 1.5 volt AAA-size batteries power the Model 9560 for approximately 
600 spot checks.

Storage:

12 months, minimum

Classifications per IEC 60601-1 / CAN/CSA-C22.2 No. 601.1/ UL 60601-1:

Degree of Protection: 

Type BF-Applied Part

Enclosure Degree of Ingress Protection:

IP3

2

Mode of Operation: 

Continuous

This product complies with ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing. 

This device is not made with natural rubber latex.

*± 1 A

rms

 represents approximately 68% of measurements.

**This information is especially useful for clinicians performing photodynamic therapy.
*** When the Model 9560 is transferred from a non-operating temperature/humidity condition, allow one hour of stabilization to operating temperature/

humidity specifications prior to use.

Accuracy Summary by Decade

Decade

Oxygen Saturation 

(A

rms

)

Low Perfusion Oxygen 

Saturation (A

rms

)

70 – 80%

±2

±2

80 – 90%

±2

±2

90 – 100%

±2

±2

70 – 100%

±2

±2

This graph shows plots of the error (SpO

2

 – SaO

2

) by SaO

2

 using 

the 9560 with a linear regression fit and upper 95% and lower 
95% limits of agreement. Each sample data point is identified by 
subject from a clinical study in non-motion conditions. 

Summary of Contents for Onyx II 9560

Page 1: ...tional hemoglobin Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following The device may not work when circulation is reduced Warm or rub th...

Page 2: ...r assistance RF Exposure For body worn operation to maintain compliance with FCC RF exposure guidelines use only accessories that contain no metallic components and provide a separation distance of 15...

Page 3: ...ch new Model 9560 When batteries are low the numeric displays flash once per second Remove batteries if the device will be stored for more than 30 days Replace low batteries as soon as possible using...

Page 4: ...the startup sequence the device begins sensing the pulse indicated by the blinking pulse quality indicator Allow the device to stabilize and observe about 4 seconds of continuous green colored pulse...

Page 5: ...as permanent damage could result 2 Dry with a soft cloth or allow to air dry Ensure that all surfaces are completely dry Testing Summary SpO2 accuracy and low perfusion testing was conducted by Nonin...

Page 6: ...imately 600 spot checks Storage 12 months minimum Classifications per IEC 60601 1 CAN CSA C22 2 No 601 1 UL 60601 1 Degree of Protection Type BF Applied Part Enclosure Degree of Ingress Protection IP3...

Page 7: ...l 9560 found to be within specifications Model 9560 is a precision electronic instrument and must be repaired by trained Nonin personnel only Any sign or evidence of opening the Model 9560 field servi...

Page 8: ...netic environment specified below The customer and or user of this device should ensure that it is used in such an environment Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV cont...

Page 9: ...20 V m 80 MHz to 2 5 GHz 20 V m where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in mete...

Reviews: