![Nonin LifeSense LS1-9R Operator'S Manual Download Page 15](http://html1.mh-extra.com/html/nonin/lifesense-ls1-9r/lifesense-ls1-9r_operators-manual_1700089015.webp)
9
Table 1: Start-up Screen Icons and Display Descriptions
No.
Name
Description
1.
Audible Alarm
Disable
Pressing this icon turns the audible alarms off.
It disables the audible alarms by setting all lower limits to 0.
2.
Audible Alarm
Enable
Pressing this icon turns the audible alarms on.
Default if no icon is chosen.
3.
≤
30kg/66lbs
Pressing this icon selects the default alarm limits for patients
weighing 30 kg (66 lbs) or less.
Only available if Audible Alarm Enable icon is chosen on previous
screen.
4.
>30kg/66lbs
Pressing this icon selects the default alarm limits for patients
weighing more than 30 kg (66 lbs).
Only available if Audible Alarm Enable icon is chosen on previous
screen.
5.
LifeSense Version
Shows LifeSense version.
If an error occurs during start-up, an error number displays here
and an alarm activates.
6.
Software Revision
Level
SW: REV
Shows the software revision level installed on the LifeSense
monitor.