16
Operator Functions
The Model 7500FO features a number of advanced options, which are intentionally more difficult
to activate. These functions are recommended only for trained operators and require multiple
button presses to prevent accidental activation.
Alarm Volume
“
adb
”
2
“
Hi
” “
0FF
” or “
Lo
” or “
Hi
”
Clear Memory
“
CLr
”
1
“
no
” “
yES
” or “
no
”
Confirm Memory Clear
“
dEL
”
1
“
no
” “
yES
” or “
no
”
Year
“
y
” “
00
”
0 to 99 by 1
Month
“
nn
” “
00
”
0 to 12 by 1
Day
“
d
” “
00
”
1 to 31 by 1
Hour
“
h
” “
00
”
0 to 23 by 1
Minutes
“
nn
” “
00
”
0 to 59 by 1
Notes:
1) Both of these menu options are part of the memory clear command; “
dEL
” will be displayed only if
“
yES
” was selected as the setting for the “
CLr
” parameter.
2) These parameters are restored when Recall Alarm Settings is set to “
yES
.” These are also the settings
displayed by Review Alarm Settings.
3) The low SpO
2
Alarm limit saved for recall cannot be lower than the current default for that alarm limit. If
it is, the default value will be used when alarm limits are restored.
Table 5: Advanced Functions
Function
Button
Instruction
Recall Previous
Alarm Limit
Settings
Press the Limits button while the unit is on. “
rCL
”
appears, indicating that previous alarm limit
settings may be recalled. To recall the settings,
press the Plus button and select “
yES
.” Press the
Limits button again to confirm.
Memory Playback
Press and hold the Plus (+) button while turning on
the Model 7500FO. This functions with the Nonin
nVISION
®
software. Select the Model 7500 option
in nVISION software.