NOISE COM, INC.
Q: inQuiry
The 'Q' command is issued by the controlling computer or terminal
to solicit a Status Report of the current state of the PNG- III.
Upon receiving the command, the PNG will respond by transmitting
a string of ten (10) 4-digit numbers, each separated by a
comma and followed by the single digit '0' or '1' to indicate
the status of Local Lock-Out. The line is terminated by a CR/LF
sequence. The first ten numbers represent: Noise_Atten,
External_Atten, Noise_Filter, External_Filter, Noise_Step,
External_Step, Noise_Time_Delay, External_Time_Delay,
Noise_Power, External_Power. The presented valued will have
validity only for those parameters whose features have been
installed in the PNG-III. Typical Format is:
'1270,1270,0001,0001,0010,0010,0000,0000,0000,0000,0crlf'
Note that values for Attenuation, Step and Time parameters are
reported without decimal points. It should be recognized that a
decimal point is appropriate between the 3rd and 4th digits of
these values.
Document M016301
Pg 24
Rev. 2 2/26/03
Local Lock
0/1=Off/On
External Power
Measure
Noise Power
Measure
External time
Delay
Noise Function
Time Delay
Ext. Attenuation
Step size
Noise Atten
Step Size
External Filter
Setting
Noise Filter
Setting
External Signal
Attenuator
Noise Attenuator
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com