NOISE COM, INC.
INITIAL PNG CONFIGURATION
The first time the PNG is used, and whenever the Remote
Communications requirements change, the 'LOCAL' Function should
be selected from the Keypad. By pressing this key as the
first in a sequence (key interpretations are sequence
dependent), the following information screen will be displayed:
The above display indicates that the PNG is set for remote
control via the GPIB (IEEE-488) interface, and that its GPIB
address is 5. The 'bar' underlining the New Setting '0000' is
a cursor, inviting the user to alter this parameter. A new
value may be entered from the keyboard, followed by 'ENTER'. If
only the 'ENTER'
key is pressed, without entering any new digits, the original
value will not be changed. When the 'ENTER' key is pressed,
the New Setting will be 'burned' into non-volatile memory,
replacing the previous value. The MAIN INFORMATION DISPLAY
will re-appear.
Permissible values for the GPIB/BAUD parameter are '0000' through
'0031', which represent GPIB addresses, and the discrete values
'0150', '0300', '0600', '1200', '2400', '4800', and '9600',
which represent Baud Rates for the optional RS- 232/485
Asynchronous Serial Interface. Selecting a valid GPIB address
will simultaneously enable the Serial interface at 9600 Baud.
The Serial ASCII character configuration is 8-Data Bits, NO-
Parity, 1-Stop Bit. Lower case characters are internally
converted to upper case by the PNG. The PNG will acknowledge each
Remote Command received by sending a Carriage Return (CR =
0D-hex or 13-decimal) character to the Serial Port.
Document M016301
Pg 12
Rev. 2 2/26/03
KEYBOARD SETUP
GPIB/BAUD :0005
NEW SETTING :0000
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com