71
CONTROL settings
Internal generator power
Level
You can set the output power level of the internal function generator to between
0 and 100%. Use the slider or the field in the upper right corner of the
Level
slider
to adjust the modulation level.
Internal generator
Offset
The offset of the internal modulation signal can be adjusted between -100 and
+100% dependent on the modulation level. For example, if the modulation level is
set to 40%, the offset can be adjusted between -60 and +60%. The sum of
modulation level and the absolute offset can be a maximum of 100%.
Integrating Update Interval
The integrating update interval sets the update interval for a software integrating
function. The function is only available on X15 systems with
Coupling
configured
as
DC
and
Range
set to
Narrow
.
Turning on wavelength modulation
To enable wavelength modulation, in the status panel lower right click the WL
mod button. The indicator next to the button turns ON green. Clicking the button
again once turns of the feature.
Figure 42
Turning on wavelength modulation
Power/Current
mode
The laser output controls can be set to either Power or Current mode as
described in
. Click the menu item shown in
to toggle the mode. The Control panel will automatically switch to the
controls for the new mode.
Figure 43
Power and current mode controls
Click to enable
modulation
wavelength
Summary of Contents for Koheras ADJUSTIK
Page 1: ...Koheras ADJUSTIK PRODUCT GUIDE Low Noise Single Frequency Laser System ...
Page 10: ...10 ...
Page 12: ...12 ...
Page 16: ...16 ...
Page 18: ...18 ...
Page 32: ...Chassis labels 32 ...
Page 44: ...44 ...
Page 52: ...Overview 52 ...
Page 78: ...78 ...
Page 82: ...Placing the laser on a table or shelf 82 ...
Page 88: ...Connecting the optical output 88 ...
Page 90: ...90 ...
Page 93: ...93 Table 17 Mechanical dimensions ...
Page 94: ...94 ...
Page 104: ...Installing CONTROL 104 ...
Page 105: ...Koheras ADJUSTIK Product Description Revision 1 0 01 2020 W 10456 ...