8 / 28
tmh5554a0600p_06
tmh555en2106p
3.
TECHNICAL DATA
Type
Rigid probe version
T
R, T
A, T
J
−
−
Flexible probe version
T
K, T
E, T
N,
T
L
−
−
Flexible, plastic-coated probe version
T
H, T
C, T
F,
T
G
−
−
Insertion length
1…4 m (3.3…13 feet)
1…50 m (3.3…164 feet)
1…50 m (3.3…164 feet)
Highest medium pressure
25 bar (2.5 MPa, 363 psi)
16 bar (1.6 MPa, 232 psi)
3 bar (0.3 MPa, 43.5 psi)
Ambient temperature
(1)
With plastic housing:
–30…+65 ºC (–22…+149 ºF)
With metal housing:
–30…+65 ºC (–22…+149 ºF)
with SAP–300 display: –20…+60 ºC (–4…+140 ºF)
Medium temperature
(1)
–40…+105 ºC (for max. 1 hour: +125 ºC)
[–40…+221 °F (for max. 1 hour: +257 °F)]
–40…+80 ºC (for max. 1 hour: +85 ºC)
[–40…+176 °F (for max. 1 hour: +185 °F)]
Accuracy
–40…–10 ºC ±2 ºC
(–40…+14 °F: ±3.6 °F);
–10…+85 ºC ±0.5 ºC (+14…+185 °F: ±0.9 °F);
+85…+125 ºC ±2 ºC (+185…+257 °F: ±3.6 °F)
Resolution (digital)
0.1 ºC (+0.18 °F)
Material of wetted parts
Stainless steel: 1.4571
Stainless steel: 1.4571 + 1.4301
Stainless steel: 1.4571
+ Antistatic PE-coated steel + 1.4301
Output
Analog 4…20
mA
Digital
4…20 mA HART
®
communication (minimal loop resistance: 250
Ω
)
Display SAP
−
300 LCD
Error indication
3.9 mA or 20.5 mA
Maximum output load
R
max
= (U
Supply
– U
Supply min.
) / 0.02 A;
Power supply (U
Supply
)
(1)
11…36 V DC (in case of HART
®
multi-drop: 10…36 V DC)
Electrical protection
Class III
Electronics housing's ingress
protection
IP67
Probe housing's ingress protection
IP68 (up to medium pressure)
IP66
Process connection
As per order codes
Electrical connection
2× M20×1.5 plastic cable gland, cable outer diameter: Ø6…12 mm (Ø.236…Ø.472"), wire cross section: max. 1.5 mm
2
(AWG16)
Housing material
Powder-coated cast aluminum (EN AC-42000), stainless steel (1.4571/Ti316) or plastic (PBT)
Weight
1.7 kg + probe: 0.6 kg/m
(3.75 lb + probe: 0.4 lb/ft)
2.9 kg + probe cable: 0.3 kg/m + weight 3 kg
(6.4 lb + probe cable: 0.2 lb/ft + weight: 6.6 lb)
2.9 kg + probe cable: 0.7 kg/m
(6.4 lb + probe cable: 0.5 lb/ft)
(1)
For explosion-proof design, see section 3.1 Ex information