Temperature simulation cycle
Parameter: P85
Menu path:
SERVICE / TEMP. SIMULATION / TIME
Default value:
60 sec
Description:
Cycle period of the simulation
Bottom value of the Temperature simulation
Parameter: P86
Menu path:
SERVICE / TEMP. SIMULATION / BOTTOM VALUE
Default value:
–10 ºC
Description:
Lowest value of the simulation
Upper value of the Temperature simulation
Parameter: P87
Menu path:
SERVICE / TEMP. SIMULATION / UPPER VALUE
Default value:
85 ºC
Description:
Highest value of the simulation
5.3.4.4
Load default values
Parameter: --
Menu path:
SERVICE / DEFAULTS / LOAD DEFAULT
Description:
This command loads all default values of the instrument.
After loading the default values, the parameters can be freely changed, but the changes do not take effect until the user exits
programming mode and returns to measurement mode. Before loading the default values, the device asks for confirmation. By loading
the default values, all customized values will be lost!
5.3.4.5
Restart
Parameter: --
Menu path:
SERVICE / RESTART
Description:
Restarts the instrument (Cold boot) (Reloads parameters from the non-volatile memory)