37
KTB/03/15.01.2020
16.
EMC (Electromagnetic Compatibility)
Decalaration
This equipment generates, uses and can radiate radio frequency (RF) energy. This
equipment may cause electromagnetic interference if it is not installed and used
in accordance with the manual.
This device has been tested for Medical Products in accordance with EN-60601-1-
2:2015 and has been determined to comply with acceptable limits. These limits
indicate that the device provides an acceptable level of protection against
electromagnetic interference (EMC) if the device is used as specified in the
manual.
This device may be affected by portable and mobile RF communication devices.
This equipment must not be stored together with other equipment.
See the following tables for more information about this device and EMC.
The paper form of the Electromagnetic Compatibility (EMC) declarations is
available free of charge within 7 days if requested.
Guidance and manufacturer’s declaration –
electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the user of
the device should assure that it is used in such an environment
Emission test
Compliance
Electromagnetic environment
–
guidance
Emissions RF
CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The device is suitable for use in all establishments including domestic and
those directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic
emissions used
for domestic
purposes.
IEC 61000-3-2
N/A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
N/A