PROCEDURE FOR CHANGING THE HEPA FILTER
(1) The precautionary steps described in this manual should be used when changing the HEPA
filter.
(2) NOTE: You must follow Local and 1 or Federal guide lines when working with
contaminated filters.
(3) Remove the eight (8) bolts holding the HEPA filter to the lid assembly, using a Phillips
screw driver and an 11/32" wrench.
(4) Carefully remove the HEPA filter, being careful not to create any airborne dust.
(5) This filter cannot be cleaned, and should be disposed of in an appropriate manner.
(6) Re-install the new HEPA filter using #8 screws, nuts, lock washers, and flat washers.