Chapter 5
Assembly
80
(5) Attach the glass plate supplied with the P-SXY64 XY
Stage so that it fits the upper plate of the stage. The
glass plate has chamfered corners including one
small chamfered corner. Align the small chamfered
corner with the right rear side of the stage (as
viewed from the front) to fit the glass plate on the
stage, and tighten the glass plate fixing screw.
11
Attach an SMZ10A Aperture Unit.
Attach the SMZ10A Aperture Unit after performing the
previous section 3.1, step 3, Attach a zooming body.
(1) Completely loosen the M4 set screw at the upper
front of the zooming body using a hex wrench
(nominal designation: 2).
(2) Align the positioning groove at the bottom of the
aperture unit with the positioning pin on the rear of
the round dovetail groove at the zooming body
upper plane so that they fit, and then tighten the
fixing screw.
(3) Attach a tube on top of the aperture unit.
Align the positioning groove with the positioning pin
of the aperture unit in the same way as the previous
section 3.1, step 4, and tighten the tube fixing screw
on the aperture unit to securely attach the tube.
SMZ
800
Securely attaching the aperture unit
Using together with a coaxial epi illuminator
When using the SMZ10A Aperture Unit together with the coaxial epi illuminator, first attach the SMZ10A Aperture
Unit, and then attach the coaxial epi illuminator on top.
12
Attach a P-IBSS2 Beam Splitter.
Attach the P-IBSS2 Beam Splitter after performing 3.1,
step 3 "Attach a zooming body.”
(1) Completely loosen the M4 set screw for fixing the
tube on top front of the zooming body using a hex
wrench (nominal designation: 2).
(2) Align the positioning groove at the bottom of the
beam splitter with the positioning pin on the rear of
the round dovetail groove on the zooming body
upper plane so that they fit, and then tighten the
fixing screw.
The port of the P-IBSS2 can be placed both on the
right and the left sides.
(3) Attach a tube on the beam splitter.
Align the positioning groove of the tube with the
positioning pin of the beam splitter to attach a tube in
the same way as the previous section 3.1, step 4.
Tighten the tube fixing screw of the beam splitter to
fix the tube in place.
P-IBSS2 port position
Securely attaching the beam splitter
Positioning pin
Positioning pin
Positioning pin
Positioning pin
Summary of Contents for SMZ1270
Page 1: ...Stereo Microscope SMZ800N SMZ1270 SMZ1270i Instruction Manual M609 E 18 5 NF 7 1 2 M609EN07 ...
Page 2: ......
Page 18: ...Chapter 1 Components 4 ...
Page 22: ...Chapter 2 Microscopy Procedures 8 ...
Page 96: ...Chapter 5 Assembly 82 ...
Page 102: ...Chapter 7 Maintenance and Storage 88 ...
Page 103: ...89 Chapter 8 Specifications ...