Chapter 4 - Technical Specification
Immunity test
IEC60601 compliance test
level
Electromagnetic environment-guidance
Electrical fast
transient/burst
IEC61000-4-4
±2 kV for power supply lines
±1 kV for input / output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC61000-4-5
±1 kV line to line
±2 kV line to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
lines
IEC61000-4-11
<5% U
T
(>95%dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
device requires continued operation during mains
interruptions, it is recommended that the device be
powered from an uninterruptible power supply.
<5% U
T
(>95% dip in U
T
)
for 5 seconds
If the user of the device requires continued operation
during mains interruptions, it is recommended that the
device be powered from an uninterruptible power supply.
Power frequency
(50/60Hz)
Magnetic field
IEC61000-4-8
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Note: UT is the a.c. mains voltage prior to the application of the test level.
Conductive RF
IEC61000-4-6
3Vrms
150kHz to 80MHz
Portable and mobile RF communications equipment
should be used no closer to any part of the device,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d=1.2√P
Page 68 of 76
Part Number: G108707/8GSE
English
Copyright 2018, Optos plc. All rights reserved.
Summary of Contents for Optos Monaco P200TE
Page 4: ...Page 4 of 76 Part Number G108707 8GSE English Copyright 2018 Optos plc All rights reserved ...
Page 10: ...Page 10 of 76 Part Number G108707 8GSE English Copyright 2018 Optos plc All rights reserved ...
Page 20: ...Page 20 of 76 Part Number G108707 8GSE English Copyright 2018 Optos plc All rights reserved ...
Page 28: ...Page 28 of 76 Part Number G108707 8GSE English Copyright 2018 Optos plc All rights reserved ...
Page 74: ...Page 74 of 76 Part Number G108707 8GSE English Copyright 2018 Optos plc All rights reserved ...
Page 76: ...Page 76 of 76 Part Number G108707 8GSE English Copyright 2018 Optos plc All rights reserved ...
Page 77: ...Hrvatski Croatian optos com ...
Page 155: ...Türkçe Turkish optos com ...
Page 233: ...Ελληνικά Greek optos com ...
Page 318: ...G108707 8GSE English Croatian Turkish Greek ...