CAUTION
Wear powder-free gloves at all times when cleaning the device including the scan head
and main mirror. If the gloves become contaminated they must be replaced immediately.
Always follow the cleaning and disinfecting agents’ manufacturer’s instructions and
precautions; for storage, method of application, protective gloves, shelf life and disposal
to help assure effective and safe use of the product.
Be careful
NOT
to damage the scan head covers when cleaning.
Do
NOT
use lint cloths, tissues, or other materials that may create dust, on or near the
scan head. Always allow cleaned areas to air-dry.
External surfaces must be cleaned of all visible contamination when they become dirty or soiled with bodily
fluids. Follow your internal clinic or hospital decontamination procedures.
Cleaning of the external surfaces of the device should be carried out on a regular basis, see
3.4.7.4 Battery replacement
CAUTION
Do
NOT
remove the scan head covers. Only Optos trained technical personnel are
permitted to service the device.
A 3V Lithium coin battery is embedded in the hardware for the device and does
NOT
require to be changed by
users. If there is a problem, please contact your Optos representative, see
3.4.7.5 Cable management
WARNING
Do
NOT
plug the scan head into a power block or powered table. The scan head must be
plugged directly into a wall socket.
To prevent electric shock:
l
Only use cables supplied by Optos.
l
Cables should
NOT
be extended or altered.
l
Mains supply must be earthed.
l
The device may only be serviced by Optos trained personnel.
l
Care must be taken never to touch exposed parts of the device while in physical
contact with a patient.
If the mains cable is damaged do not use the system and contact your Optos
representative, see
CAUTION
Ordinary equipment (IP2X), without protection against ingress of water.
You must regularly inspect the equipment. If there is any damage please contact your
Optos representative, see
Ensure the mains socket is easily accessible.
Part Number: G108707/8GSE
Page 59 of 76
Copyright 2018, Optos plc. All rights reserved.
English
Summary of Contents for Optos Monaco P200TE
Page 4: ...Page 4 of 76 Part Number G108707 8GSE English Copyright 2018 Optos plc All rights reserved ...
Page 10: ...Page 10 of 76 Part Number G108707 8GSE English Copyright 2018 Optos plc All rights reserved ...
Page 20: ...Page 20 of 76 Part Number G108707 8GSE English Copyright 2018 Optos plc All rights reserved ...
Page 28: ...Page 28 of 76 Part Number G108707 8GSE English Copyright 2018 Optos plc All rights reserved ...
Page 74: ...Page 74 of 76 Part Number G108707 8GSE English Copyright 2018 Optos plc All rights reserved ...
Page 76: ...Page 76 of 76 Part Number G108707 8GSE English Copyright 2018 Optos plc All rights reserved ...
Page 77: ...Hrvatski Croatian optos com ...
Page 155: ...Türkçe Turkish optos com ...
Page 233: ...Ελληνικά Greek optos com ...
Page 318: ...G108707 8GSE English Croatian Turkish Greek ...