IV. Assembly
53
7
Attaching the Polarizing Intermediate Tube
1
Loosen sufficiently the clamp screw for the
intermediate tube on the microscope arm (or on
the illuminator).
Use the tool provided with the microscope to
loosen the clamp screw for the epi illuminator.
2
Fit the circular dovetail of the polarizing
intermediate tube into the circular dovetail
groove of the microscope arm (or of the epi
illuminator).
When fitting, insert the positioning pin on the
polarizing intermediate tube into the receiving
groove on the arm (or the epi illuminator).
3
Secure the polarizing intermediate tube by
tightening the clamp screw.
Remove any looseness between the positioning pin and groove by pushing in the polarizing
intermediate tube while rotating in the clockwise direction.
8
Attaching the Eyepiece Tube
Loosen the eyepiece tube clamp screw on the
intermediate tube with the provided tool sufficiently.
Attach the eyepiece tube onto the mount part on the
illuminator and fix it with eyepiece tube clamp screw
using the hexagonal screwdriver.
Caution to remove the eyepiece tube
Hold the eyepiece tube by hand when loosing the
clamp screw to prevent a sudden disconnection and
falling.
9
Attaching Eyepieces
Attach eyepieces of the same magnification and the same field number.
There are positioning protrusions on the binocular part sleeve of the eyepiece tube. Align the
notches of the eyepieces with the protrusions on the sleeve and slide the eyepieces into the eyepiece
sockets.
0
Clamp screw
Polarizing
intermediate
tube
0
BINO
PHOTO
PHOTO
BINO
&
Clamp screw
Trinocular
eyepiece tube