![Nikon Eclipse LV100POL Instructions Manual Download Page 17](http://html1.mh-extra.com/html/nikon/eclipse-lv100pol/eclipse-lv100pol_instructions-manual_1691841017.webp)
15
II. Microscopy
18
Rotate the nosepiece to locate the desired
magnification objective into the optical
path. Then, perform the microscopy.
•
Each time you change objectives, the field
diaphragm and the condenser aperture
diaphragm must be adjusted.
- The field aperture should be adjusted so that
its image circumscribes the view field.
- And the aperture diaphragm should be 70% to
80% of the numerical aperture of the
objective.
See Pages 31 and 32.
•
Focus on the specimen again using the fine focus
knob or the coarse focus knob.
See Page 25.
•
The top lens of the swing-out condenser must be
operated depending on the magnification of the
objective to be used.
See Page 29.
•
To use an oil immersion type objective or a
water immersion type objective, perform the oil
immersion or water immersion.
See Pages 34 and 35.
19
When the microscopy ends, turn off the power switch.
ND8
NCB
F.S.
0.4
0.2
0.8
0.6
JAPAN
Achr
0.90
0
BINO
PHOTO
PH
OTO
BINO
&
18
Switch to any
desired objective.
18
Adjust the
condenser
aperture
diaphragm.
18
Adjust the field
diaphragm.
18
Readjust
the focus.