Troubleshooting Tables
VIII
If difficulties should be encountered in the course of operation, please recheck the symptoms,
referring to the tables below, before contacting your nearest Nikon representative. If the
problem is still not solved after referring to Troubleshooting Tables, please contact your nearest
Nikon representative.
1
Optical
Troubles Causes
Corrective
Measures
Revolving nosepiece not in
click-stop position (objective not
centered in the optical path).
Revolve to click-stop position
(swing the objective correctly
into the optical path).
Condenser position too low.
Position the condenser slightly
lower than the upper limit.
(P. 38)
Condenser not installed
correctly.
Install correctly. (P. 38)
Field lens unit not installed
correctly.
Install correctly. (P. 40)
Lamp not installed correctly.
Install correctly. (P. 39)
Dirt or dust on the lens
(condenser, objective, field lens,
eyepiece, specimen)
Clean the lens. (P. 49)
Optical path selection lever on
the trinocular eyepiece tube is
not in click position.
Optical path selection lever on
the trinocular eyepiece tube is
not set for 100% binocular part.
Set the lever correctly for 100%
binocular part.
Filter is out of position.
Set correctly on the field lens.
The top lens of the swing-out
condenser is not set in correct
position.
Swing in or out till the limit.
Bertland lens in the optical path.
Remove it.
P-CL, P-CS or P-CQ sliders not
set correctly.
Set them in correct position.
Darkness at the periphery,
no viewfield seen, or uneven
viewfield brightness.
Analyzer slider in intermediate
position.
Set correctly.