![Nikon ECLIPSE E200 POL Instructions Manual Download Page 23](http://html1.mh-extra.com/html/nikon/eclipse-e200-pol/eclipse-e200-pol_instructions-manual_1691829023.webp)
IV
Microscopy (Detailed Procedure)
21
IV
8
Aperture Diaphragm Adjustment
z
Orthoscopic Microscopy
The aperture diaphragm is important because it
is related to the resolution, contrast, depth of
focus and brightness of the optical image.
Turning the condenser aperture diaphragm ring
changes the size of the aperture diaphragm.
As the aperture diaphragm is stopped down,
resolution and brightness are reduced while
contrast and depth of focus are increased.
Conversely, as the aperture diaphragm is
opened, resolution and brightness are increased
while contrast and depth of focus are reduced.
It is not possible to adjust one pair of
characteristics without affecting the other.
Generally, a satisfactory image with appropriate
contrast can be obtained with an aperture
setting that is 70% to 80% of the numerical
aperture of the objective. The numerical
aperture is indicated on the barrel of each
objective.
An indication of 40x / 0.65 means that the
magnification is 40x and the numerical
aperture is 0.65.
If the aperture diaphragm is stopped down too
far, the resolution is reduced; therefore, except
when viewing a nearly transparent specimen,
we do not recommend stopping down the
aperture to less than 60% of the numerical
aperture of the objective.
z
Adjusting the size of the aperture diaphragm according to the condenser scale
Since the condenser scale indicates the numerical aperture, adjust the aperture
diaphragm ring according to the scale. (Normally, the index on the aperture diaphragm
ring should be aligned with the scale line corresponding to 70% to 80% of the numerical
aperture of the objective.)
z
Adjusting the size of the aperture diaphragm using the Bertrand lens
Insert the Bertrand lens into the optical path (by placing in position “B”). Turn the
diaphragm control ring to stop down the aperture diaphragm to its minimum setting.
Turn the Bertrand lens focus ring to focus on the aperture diaphragm. Turn the
diaphragm control ring to adjust the aperture diaphragm. (This is normally adjusted to
70-80% of the field of view.)
Numerical
aperture
0.65
Adequate size of the
condenser aperture diaphragm
=Objective N.A. x0.7
=0.65x0.7=0.455
Set to 0.455