Nidek Medical Mark 5 Nuvo OCSI Lite 3 User Manual Download Page 8

2010-8403CE Rev C

8

September 2013

5.4.3 Indicators (continued)

The red indicator warns of a process
fault. One event that can cause the red
indicator to be illuminated is low oxy-
gen concentration. The low oxygen con-
centration red indicator will light when
oxygen concentration falls below a pre-
determined set point. Another event that
will cause the red indicator to light is a
blocked cannula. In this case the green
indicator and red indicator will be illumi-
nated simultaneously.

5. 5. OCSI (

oxygen concentration status

indication module)

function

5.5.1. Operating principle

The oxygen monitor (Item 2 Fig 2.1) is an
electronic module capable of checking the
effective oxygen concentration supplied by
the

Mark 5 Nuvo Lite

concentrator.

The oxygen monitor measures the concentra-
tion and activates an audible and visual alarm if
it falls below the alarm set point percentage.

(Refer to Section 5.4 for information on the
operation of the indicators and alarms for the
OCSI function)

In compliance with the ISO 8359 standard.

The fl ow supplied is equal to the selected fl ow, 

accurate to within ± 10 % or 200 ml/min,
whichever is the larger.

Oxygen Concentration

• at 2 l/min: >90%.
• at 3 l/min: 90%. (+6.5%/-3%)

(Values at 21

o

C and at one atmosphere pres-

sure).

Maximum fl ow: 3 l/min.

The variation of the maximum fl ow does not 

exceed ± 10 % of the indicated value when a
back pressure of 7 kPa (1 psig) is applied to
the output of the device. The maximum outlet
pressure is 50 kPa (7 psig).

5.5.4 Maintenance of the Device Alarms

No special maintenance is required. The alarm
set-point is factory set and the setting cannot be
adjusted. All models are set at 84%.

The equipment supplier verifi es that the device 

is still operating correctly when the routine
checks are performed.

5.6. Technical characteristics

Dimensions: L x W x H: 36x23x58.5 cm (14
x 9 x 23 in.)
Caster diameter: 3.8 cm (1.5 in.).

Tilt  angle  (transport  with  humidifi er  fi tted) 

30

o

.

Weight: 14.5 kg. (32 lbs) varies by model.
Noise level conforms to ISO 8359 Standards.

Flow values

8 position fl ow valve 0.125-3 liters/minute.

(Some models may have other values.)

Accuracy of fl ow supplied

(Fig. 5.1)

Red Indicator

Green Indicator

Summary of Contents for Mark 5 Nuvo OCSI Lite 3

Page 1: ...gen concentration warning light GLOSSARY OF SYMBOLS Original language is English GLOSSARY OF SYMBOLS 1 GENERAL SAFETY GUIDELINES 2 1 Unpacking and Packaging 3 1 1 Method for disposing of waste 3 1 2 M...

Page 2: ...ing to quali ed service personnel GENERAL SAFETY GUIDELINES Only persons who have read and understood this entire manual should be allowed to operate the MARK 5 Nuvo Lite TheCAUTIONSbelowindicateapote...

Page 3: ...detected please contact your equipment provider Operating environmental condi tion guidelines are discussed later in another section of this User s Guide 1 1 METHOD FOR WASTE DISPOSAL All waste from t...

Page 4: ...r the humidi er bottle manufacturer s recommendation Re attach the lid to the bottle and connect to the oxygen concentrator Place the bottle on the concentrator and secure with the attachment straps a...

Page 5: ...Adjust the nasal cannula to suit your face see Fig 3 4 1 Filter Silencer 2 Cabinet lter 3 Ventilation grill Note Shown with grate removed 115V Fig 3 3 230V Fig 4 1 3 2 Turning off device Attheendofth...

Page 6: ...cuit oxygentherapynasalcannula etc must be changed 4 3 Maintenance No special maintenance needs to be carried out by the patient Your equipment supplier performs periodic maintenance operations to ass...

Page 7: ...rms No voltage detection In the event of a loss of mains power an inter mittent audible alarm is activated and the green light turns off Test alarm by actuating the I O ON OFF switch when the power co...

Page 8: ...he OCSI function In compliance with the ISO 8359 standard The ow supplied is equal to the selected ow accurate to within 10 or 200 ml min whichever is the larger Oxygen Concentration at 2 l min 90 at...

Page 9: ...repairable Electrical power supply Rating 230V 50Hz Average Power 180 W avg Protection Class Class II Mains Protection 5A Filters At the rear of the device a cabinet air lter At the compressor input a...

Page 10: ...blockage Fan is not working Tube disconnected or humidi er cap is not tight Cannula tubing is restricted Observations The I O ON OFF button is in the I ON position but the device does not operate The...

Page 11: ...__________________ ___________________________________________________________________ ___________________________________________________________________ _____________________________________________...

Page 12: ...Drive Birmingham Alabama 35217 U S A Tel 205 856 7200 Fax 205 856 0533 info nidekmedical com www nidekmedical com EU Representative mdi Europa GmbH Langenhagener Str 71 30855 Hannover Langenhagen Ger...

Reviews: