Nidek Medical Mark 5 Nuvo OCSI Lite 3 User Manual Download Page 2

2010-8403CE Rev C

2

September 2013

.

The WARNINGS below indicate a potential

hazardous situation. If conditions are not

avoided a situation could occur that results

in serious injury or death.

Oxygen  is  not  a  fl ammable  gas,  but  it 

accelerates the combustion of materials. Do

not use in explosive atmosphere.To avoid risk

of fi re and explosion the concentrator should 

be kept away from Flames, Heat sources,

Incandescent sources, Smoking Materials,

Matches, Oil, Grease, Solvents, Aerosols,

etc. Do not allow oxygen to accumulate on

upholstery or other fabric such as bedding or

personal clothing. If concentrator is operating

while not connected to patient, position

cannula  so that the gas fl ow is diluted in the 

ambient air.

Improper patient connection to and use of

the cannula may result in injury including

strangulation.

Use of other accessories not described in this

User's Guide are not recommended. Patient

benefi t may be diminished.

No modifi cation to the equipment is allowed. 

To do so may affect patient benefi t.

Contraindications; Those who continue to

smoke (because of the increased fi re risk 

and the probability that the poorer prog-

nosis by smoking will offset the treatment

benefi t).

Device must have power to operate. In the

event of power loss and for continued op-

eration a backup source is recommended.

DO NOT disassemble due to danger of

electrical shock. Refer servicing to quali-

fi ed service personnel.

GENERAL SAFETY GUIDELINES

Only persons who have read and understood this entire manual

should be allowed to operate the

MARK 5 Nuvo Lite

The CAUTIONS below indicate a potentially

hazardous situation. If conditions are not

avoided a situation could occur that results in

property damage or minor injury or both

.

Use the power cord provided, and check that

the electrical characteristics of the power

socket used match those indicated on the

manufacturer’s plate on the rear panel of

the device.

We recommend against the use of extension

cords and adapters, as they are potential

sources of sparks and fi re.

The

Mark 5 Nuvo Lite

has an audible alarm

to warn the user of problems. In order that

the alarm may be heard, the maximum dis-

tance that the user can move away from it

must be determined to suit the surrounding

noise level.

The

Mark 5 Nuvo Lite

must only be used

for oxygen therapy and only on a medical

prescription. The indicated daily duration

and fl ow must be followed, otherwise it may 

present a risk to the health of the patient.

Do not use in a specifically magnetic

environment (MRI, X-ray, etc.). May cause

device malfunction.

This unit may be equipped with a polarized

plug. That is one blade wider than the other.

If  it  does  not  fi t  into  the  outlet,  reverse 

the  plug.  If  it  still  does  not  fi t,  contact  a 

qualifi ed  electrician.    Do  not  defeat  this 

safety feature.

Note: Medical Device Regulations require

users and service providers to report to the

manufacturer any incident that could, if
repeated, result in injury to any person.

Summary of Contents for Mark 5 Nuvo OCSI Lite 3

Page 1: ...gen concentration warning light GLOSSARY OF SYMBOLS Original language is English GLOSSARY OF SYMBOLS 1 GENERAL SAFETY GUIDELINES 2 1 Unpacking and Packaging 3 1 1 Method for disposing of waste 3 1 2 M...

Page 2: ...ing to quali ed service personnel GENERAL SAFETY GUIDELINES Only persons who have read and understood this entire manual should be allowed to operate the MARK 5 Nuvo Lite TheCAUTIONSbelowindicateapote...

Page 3: ...detected please contact your equipment provider Operating environmental condi tion guidelines are discussed later in another section of this User s Guide 1 1 METHOD FOR WASTE DISPOSAL All waste from t...

Page 4: ...r the humidi er bottle manufacturer s recommendation Re attach the lid to the bottle and connect to the oxygen concentrator Place the bottle on the concentrator and secure with the attachment straps a...

Page 5: ...Adjust the nasal cannula to suit your face see Fig 3 4 1 Filter Silencer 2 Cabinet lter 3 Ventilation grill Note Shown with grate removed 115V Fig 3 3 230V Fig 4 1 3 2 Turning off device Attheendofth...

Page 6: ...cuit oxygentherapynasalcannula etc must be changed 4 3 Maintenance No special maintenance needs to be carried out by the patient Your equipment supplier performs periodic maintenance operations to ass...

Page 7: ...rms No voltage detection In the event of a loss of mains power an inter mittent audible alarm is activated and the green light turns off Test alarm by actuating the I O ON OFF switch when the power co...

Page 8: ...he OCSI function In compliance with the ISO 8359 standard The ow supplied is equal to the selected ow accurate to within 10 or 200 ml min whichever is the larger Oxygen Concentration at 2 l min 90 at...

Page 9: ...repairable Electrical power supply Rating 230V 50Hz Average Power 180 W avg Protection Class Class II Mains Protection 5A Filters At the rear of the device a cabinet air lter At the compressor input a...

Page 10: ...blockage Fan is not working Tube disconnected or humidi er cap is not tight Cannula tubing is restricted Observations The I O ON OFF button is in the I ON position but the device does not operate The...

Page 11: ...__________________ ___________________________________________________________________ ___________________________________________________________________ _____________________________________________...

Page 12: ...Drive Birmingham Alabama 35217 U S A Tel 205 856 7200 Fax 205 856 0533 info nidekmedical com www nidekmedical com EU Representative mdi Europa GmbH Langenhagener Str 71 30855 Hannover Langenhagen Ger...