Nidek Medical Mark 5 Nuvo OCSI Lite 3 User Manual Download Page 7

September 2013

2010-8403CE Rev C

7

5.2. Materials in direct or indirect contact

with the patient

Concentrator casing ................................ABS
Power Cord ........................................... PVC
Cabinet Air Filter ..............................Polyester

I

/0

(On/Off) switch................................Nylon

Casters....................................................Nylon
Flow adjustment knob..............................ABS

Oxygen product outlet ...................Aluminum

Printed labels.............................Polycarbonate

Pipe/Tubing..........................Aluminium,PVC,

polyurethane or silicone

Humidifi er.................................Polypropylene

 Humidifi er tube........................................PVC

Filter .........................................Polypropylene

5.3. Operating principle

The  compressor  sends  fi ltered  room  air  to  a    

solenoid valve, which allows compressed air to
pass to the column in production. The columns
contain a molecular sieve, whose function is
to adsorb the nitrogen and thus allow oxygen
to pass. The oxygen enriched product is then
directed through a pressure reducing valve to

the adjustable fl ow valve and continuing to the 

oxygen product outlet fi tting.    

During this time, the column which is being
"regenerated" is connected to the ambient air

and fl ow of oxygen enriched product is passed 

through it (from the column "in production"). In
this way, when one column is in production, the
other is in a nitrogen desorption or "regenera-

tion" phase. The oxygen enriched product fi nally 

passes through a fi nal product fi lter located prior 

to the oxygen outlet fi tting.

5.4. Alarms - Safety devices - Indications

5.4.1. Alarms

No voltage detection

In the event of a loss of mains power, an inter-
mittent audible alarm is activated and the green
light turns off. Test alarm by actuating the

I

/

O

(ON/OFF) switch when the power cord is not
plugged into the wall outlet.

7

Process fault

In the case of a process fault, a visible and au-
dible alarm is activated (continuous red light or
lighted alarm and audible alarm).

Oxygen Concentration

If the oxygen concentration level falls below the
required range the red light comes on and the
green light goes out. After a 15 minute delay
the audible alarm will sound.

Blocked Cannula

If the oxygen fl ow is blocked within the can-

nula and remains so for 5 seconds the red
indicator in addition to the green indicator
will be illuminated. Also, an audible alarm
will sound.

5.4.2. Safety devices

Compressor motor

Thermal safety is ensured by a thermal switch
situated in the motor winding (145 ± 5

o

C).

Electrical protection

A 5A circuit breaker is incorporated into the
front cabinet of all models.

Class II devices with insulated casings
(EN60601-1 standard)

Safety valve

This is fi tted on the compressor outlet and is 

calibrated to 2.7 bar (40 psig).

5.4.3 Indicators

The green indicator light (Fig.5.1) indicates
that power is applied to the device. When

fi rst turned on the indicator will fl ash until 

correct oxygen concentration is achieved.
At that time the green indicator will remain
illuminated and the device is ready to provide
oxygen enriched air to the patient.

Summary of Contents for Mark 5 Nuvo OCSI Lite 3

Page 1: ...gen concentration warning light GLOSSARY OF SYMBOLS Original language is English GLOSSARY OF SYMBOLS 1 GENERAL SAFETY GUIDELINES 2 1 Unpacking and Packaging 3 1 1 Method for disposing of waste 3 1 2 M...

Page 2: ...ing to quali ed service personnel GENERAL SAFETY GUIDELINES Only persons who have read and understood this entire manual should be allowed to operate the MARK 5 Nuvo Lite TheCAUTIONSbelowindicateapote...

Page 3: ...detected please contact your equipment provider Operating environmental condi tion guidelines are discussed later in another section of this User s Guide 1 1 METHOD FOR WASTE DISPOSAL All waste from t...

Page 4: ...r the humidi er bottle manufacturer s recommendation Re attach the lid to the bottle and connect to the oxygen concentrator Place the bottle on the concentrator and secure with the attachment straps a...

Page 5: ...Adjust the nasal cannula to suit your face see Fig 3 4 1 Filter Silencer 2 Cabinet lter 3 Ventilation grill Note Shown with grate removed 115V Fig 3 3 230V Fig 4 1 3 2 Turning off device Attheendofth...

Page 6: ...cuit oxygentherapynasalcannula etc must be changed 4 3 Maintenance No special maintenance needs to be carried out by the patient Your equipment supplier performs periodic maintenance operations to ass...

Page 7: ...rms No voltage detection In the event of a loss of mains power an inter mittent audible alarm is activated and the green light turns off Test alarm by actuating the I O ON OFF switch when the power co...

Page 8: ...he OCSI function In compliance with the ISO 8359 standard The ow supplied is equal to the selected ow accurate to within 10 or 200 ml min whichever is the larger Oxygen Concentration at 2 l min 90 at...

Page 9: ...repairable Electrical power supply Rating 230V 50Hz Average Power 180 W avg Protection Class Class II Mains Protection 5A Filters At the rear of the device a cabinet air lter At the compressor input a...

Page 10: ...blockage Fan is not working Tube disconnected or humidi er cap is not tight Cannula tubing is restricted Observations The I O ON OFF button is in the I ON position but the device does not operate The...

Page 11: ...__________________ ___________________________________________________________________ ___________________________________________________________________ _____________________________________________...

Page 12: ...Drive Birmingham Alabama 35217 U S A Tel 205 856 7200 Fax 205 856 0533 info nidekmedical com www nidekmedical com EU Representative mdi Europa GmbH Langenhagener Str 71 30855 Hannover Langenhagen Ger...

Reviews: