Nidek Medical Mark 5 Nuvo 10 User Manual Download Page 5

                                   November 2013

2010-2324CE Rev A

5

 

IV.3. Operating principle

The compressor sends filtered ambient air to an electronic 

valving system, which allows compressed air to pass to the 

column in production. The columns contain a molecular 

sieve, whose function is to adsorb the nitrogen and thus allow 

oxygen to pass.

The oxygen enriched product is then directed to a pressure 

reducing valve through the adjustable flow meter to the 

oxygen outlet fitting.

During this time, the column which is being "regenerated" 

is connected to the ambient air and flow of oxygen enriched 

product is passed through it (from the column "in production"). 

In this way, when one column is in production, the other is in 

a nitrogen desorption or "regeneration" phase. The oxygen 

enriched product finally passes through a bacterial filter located 

prior to the oxygen outlet fitting.

IV.5. ALARMS INDICATORS                                                              

IV5.1. OCSI (

oxygen concentration status indicator)

 if 

equipped                                                                                           

                                                                                                                                                                         

The oxygen monitor (6) is an electronic module capable of 

checking the effective oxygen concentration supplied by the 

Mark 5 Nuvo MB

 concentrator.

The oxygen monitor measures the concentration and activa

-

tes an audible and visual alarm if it falls below the alarm set 

point percentage. 

When the 

Mark 5 Nuvo MB

 is started, the indicator lights 

operates as follows:

IV.5.2. Green indicator 

The green indicator light indicates that power is applied 

to the concentrator and that it is ready to provide oxygen 

enriched air to the patient.  To be lighted, it is necessary that 

the concentrator power plug be inserted into the wall outlet, 

that the I/O (On/Off) switch be actuated and  that the oxygen 

concentration has reached the alarm set-point (OCSI equipped 

). On initioal power up it will flash green until the unit has 

reached normal operating conditions.

IV.6. Technical characteristics 

Dimensions: L x W x H: 394 x 396 x 706 mm (15.5 x 15.6 

x 27.8 in.) (Nuvo 10 H: 760mm (29.8))

Caster diameter: 50 mm (2.0 in.).

Tilt angle (transport with humidifier fitted): 70

o

.

Weight:  24 kg /54 lbs

Noise level within ISO 8359 guidelines

Flow values:

Continuously Adjustable Flowmeter:  1 to 5, 2 to 8, or 2 to 

10 (depending on model)

IV.5.3. Red indicator 

IV.5.4. Maintenance of the device alarms

• No special maintenance is required. The alarm set-point 

is factory set and the setting cannot be adjusted. All OCSI 

models are set at 84%.

•  The  equipment  supplier  checks  that  the  device  is  still 

operating correctly when the routine checks are performed 

on the 

Mark 5 Nuvo MB

.

IV.4. Alarms - Safety devices

IV.4.1. Alarms

• 

No voltage detection:

In the event of a loss of mains power, a intermittent audible 

alarm is activated and the green light turns off.  Test alarm by 

actuating the I/0 (On/Off) switch when the power cord is not 

plugged into the wall receptacle. 

• 

Process fault:

In the case of a process fault, a visible and audible alarm is 

activated (continuous red light or lighted alarm and audible 

alarm,  see p. 7)

.

• 

No Oxygen Flow Alarm 

(Optional):

If supplied the Nuvo MB Oxygen Concentrator has a No 

Flow Alarm.  This device provides an continuous audible 

alarm in the event the Oxygen Flow to patient becomes 

blocked.

IV.4.2. Safety devices

• 

Compressor motor:

 

Thermal safety is ensured by a thermal switch situated in the 

stator winding (145 ± 5 

o

C).

• 

Electrical protection

 of the 

Mark 5 Nuvo MB:

 

A 5A circuit breaker is incorporated into the front cabinet of all 

230V models, except NUVO 10 which is 10A. A 10 A circuit 

breaker is included with 115V models.

•  Class II devices with insulated castings (EN60601-1   

 standard)

• 

Safety valve:

This is fitted on the compressor outlet and is calibrated to 3.4 

bar (50 psig).

The red indicator light is used to warn the patient of a system 

fault.  The  two events that can cause the red indicator to be 

lighted are low oxygen concentration or abnormal pressures.

For OCSI units the low oxygen concentration warning will 

light when the oxygen concentration set point level is not 

reached.  When the  red indicator light is lighted for 15 minutes 

(±2 minutes), a continuous audible alarm is activated. Call 

the equipment supplier to service the device.

 

Summary of Contents for Mark 5 Nuvo 10

Page 1: ...3 III CLEANING MAINTENANCE 4 III 1 Cleaning 4 III 2 Everyday disinfection 4 III 3 Maintenance 4 IV USEFUL INFORMATION 4 IV 1 Accessories and spare parts 4 IV 2 Materials in direct indirect contact wi...

Page 2: ...r installer or distributor are not considered to be responsible themselves for the consequences on the safety reliability and characteristics of a device unless Theassembly fitting extensions adjustme...

Page 3: ...of the correct voltage and frequency as defined on the manufacturer s technical label 8 f Press the power switch I 0 to the ON position I The green indicator will light when the oxygen concentration e...

Page 4: ...ons The cabinet air filter should be washed or replaced The entire oxygen administration circuit oxygen therapy nasal cannulas etc must be changed III 3 Maintenance No special maintenance needs to be...

Page 5: ...angle transport with humidifier fitted 70o Weight 24 kg 54 lbs Noise level within ISO 8359 guidelines Flow values Continuously Adjustable Flowmeter 1 to 5 2 to 8 or 2 to 10 depending on model IV 5 3...

Page 6: ...the maximum recommended flow does not exceed 10 of the indicated value when a back pressure of 6 9 kPa 1 psig is applied to the output of the device The maximum outlet pressure is 7psig for 5 l min 15...

Page 7: ...vice provider for any updates to recommended schedules The work must be carried out by suitably trained technicians certified by the manufacturer Use original spare parts only see Pg 8 Uponrequest the...

Page 8: ...in the I ON position but the green indicator is not lighted Faulty indicator Contact your equipment supplier The I 0 ON OFF button is in the I ON position but there is no flow The audible alarm sound...

Page 9: ...November 2013 2010 2324CE Rev A 9 V EMC Electromagnetic Statements...

Page 10: ...010 2324CE Rev A 10 November 2013 NOTES...

Page 11: ...November 2013 2010 2324CE Rev A 11 NOTES...

Page 12: ...alley East Industrial Drive Birmingham Alabama 35217 U S A Tel 205 856 7200 Fax 205 856 0533 EU Representative mdi Europa GmbH Langenhagener Str 71 30855 Hannover Langenhagen Germany Tel 49 511 39 08...

Reviews: